Use of electronic calibration units for vector-network-analyzer verification

机译:使用电子校准单元进行矢量网络分析仪验证

获取原文

摘要

We present measurements demonstrating that electronic calibration units are stable enough to be used in place of mechanical vector-network-analyzer verification artifacts. This enables a new fully automated approach to verifying microwave vector-network-analyzer calibrations with a single computer-controlled electronic verification artifact. The new system presents verification results in easy-to-understand performance metrics that, unlike those derived from measurements of mechanical verification artifacts, are independent of the actual artifacts employed.
机译:我们提供的测量结果表明,电子校准单元足够稳定,可以代替机械矢量网络分析仪验证工件使用。这样就可以采用一种新型的全自动方法,用一个计算机控制的电子验证工件来验证微波矢量网络分析仪的校准。新系统以易于理解的性能指标呈现验证结果,该指标与从机械验证工件的测量得出的结果不同,它们独立于所采用的实际工件。