Design and Test Results of a Front-end ASIC for Radiation Detectors

机译:辐射探测器前端ASIC的设计和测试结果

获取原文

摘要

A front-end ASIC for semiconductor radiation detectors is presented. It is composed of a Charge Sensitive Amplifier (CSA),a pulse shaper,and a Peak Detect and Hold (PDH) circuit. Poly-resistor is used as source degeneration component to reduce the noise of current source in the CSA. The ASIC has been designed in a 0.5um CMOS DPTM technology and tested with Vengy 93000. The gain (PDH excluded) is 78.5mV/fC and the Equivalent Noise Charge (ENC) with detector disconnected is 800-900e. The power dissipation without the output buffer is about 2.6mW.
机译:提出了一种用于半导体辐射探测器的前端ASIC。它由电荷敏感放大器(CSA),脉冲整形器和峰值检测与保持(PDH)电路组成。多晶硅电阻器用作源极退化组件,以降低CSA中电流源的噪声。 ASIC采用0.5um CMOS DPTM技术进行设计,并通过Vengy 93000进行了测试。增益(不包括PDH)为78.5mV / fC,断开检测器时的等效噪声电荷(ENC)为800-900e。不带输出缓冲器的功耗约为2.6mW。