首页> 外文会议>Biomedical Engineering >IMPLANTABLE DEFIBRILLATORS IMMUNITY TO LOW FREQUENCY LED ELECTROMAGNETIC DISTURBANCES
【24h】

IMPLANTABLE DEFIBRILLATORS IMMUNITY TO LOW FREQUENCY LED ELECTROMAGNETIC DISTURBANCES

机译:植入式除颤器不受低频LED电磁干扰的影响

获取原文

摘要

This study intends to characterize the immunity of implantable cardioverters-defibrillators to very low frequencies electromagnetic perturbations. For that, we have set up an in-vitro evaluation protocol of the susceptibility of defibrillators to electromagnetic led-disturbances. The tested device is placed in a phantom containing gelatine to which the electromagnetic properties are close to those of human biological tissues, and we apply a configurable lure cardiac signal in order to be close to real conditions of implantation. We thus have made evident frequencies and voltage levels of led disturbing signals which mislead the defibrillator. The results obtained show that the susceptibility of the defibrillator tested increases with the voltage level of the disturbing signal and decreases with the frequency. It comes out from this report that the particulary 50 Hz frequency is very disturbing and that there exists a real EMF exposure risk for active medical implanted patients.
机译:这项研究旨在表征植入式心脏复律除颤器对非常低频率的电磁扰动的抗扰性。为此,我们建立了体外除颤器对电磁led干扰敏感性的评估协议。被测试的设备放置在含有明胶的幻影中,明胶的电磁性能与人体生物组织的电磁性能相近,并且我们施加可配置的诱饵心脏信号以接近实际植入条件。因此,我们已经得出了导致误导除颤器的LED干扰信号的明显频率和电压水平。获得的结果表明,所测试的除颤器的磁化率随干扰信号的电压电平而增加,而随频率而降低。从该报告中得出的结论是,特别是50 Hz的频率非常令人不安,并且对于活跃的医疗植入患者,确实存在EMF暴露的风险。