首页> 外文会议>ASME international mechanical engineering congress and exposition >METAL RELEASE FROM TI-6AL-4V POSTERIOR CERVICAL SPINE PLATES: A COMPUTATIONAL AND EXPERIMENTAL STUDY IN THE CANINE MODEL
【24h】

METAL RELEASE FROM TI-6AL-4V POSTERIOR CERVICAL SPINE PLATES: A COMPUTATIONAL AND EXPERIMENTAL STUDY IN THE CANINE MODEL

机译:Ti-6Al-4V后颈椎板的金属释放:犬模型的计算与实验研究

获取原文

摘要

The use of Ti-6A1-4V for spinal implants has increased due to its advantageous mechanical properties, biocompatibility, corrosion resistance, and compatibility with MRI procedures. The release of metal ions or particulates, which can result from mechanical loads imposed on the implant, can affect the degree of biocompatibility of spinal implants, possibly influencing their clinical performance. In this project, the titanium metal release from Ti-6Al-4V posterior cervical spine plates was examined using a computational and experimental approach in the canine model. The in vivo titanium metal release from plates implanted in four canines at the C4-C5 level was determined using atomic emission spectroscopy techniques. The highest titanium levels (>100 ppm dry weight) were found in tissue samples closest to the screw-plate interfaces. The retrieved implants showed the highest amounts of surface damage at the screw-plate interfaces. The histological evaluation of tissues showed a normal fibrous response with the presence of titanium particulates.
机译:由于其有利的机械性能,生物相容性,耐腐蚀性和与MRI程序的相容性,使用Ti-6A1-4V的使用增加了脊髓植入物。金属离子或颗粒的释放可以由植入物施加的机械负荷导致,可能会影响脊髓植入物的生物相容性,可能影响其临床表现。在该项目中,在犬模型中使用计算和实验方法检查来自Ti-6AL-4V后颈椎板的钛金属释放。使用原子发射光谱技术测定从C4-C5水平植入四个犬的平板中的体内钛金属释放。在最接近螺纹板接口的组织样品中发现了最高钛水平(> 100ppm干重)。检索植入物在螺杆板接口处显示出最高量的表面损坏。组织的组织学评价显示出常规纤维反应,钛颗粒的存在。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号