SURVEY OF IMAGE QUALITY AND RADIOGRAPHIC TECHNIQUE OF PEDIATRIC CHEST EXAMINATIONS PERFORMED IN BRAZIL

机译:巴西小儿胸腔检查的图像质量和射线照相技术调查

获取原文

摘要

This work presents the results of a survey of Entrance Surface Air Kerma (Ke) values, image quality and radiographic exposure parameters used in pediatric chest examinations performed in Brazil. The study is part of the activities of the IAEA Regional Project RLA/9/057 whose objective is to optimize the radiological protection of patients in diagnostic and interventional radiology, nuclear medicine and radiotherapy. The survey was performed in four hospitals in Brazil. The study group consisted of 420 pediatric patients from two age groups (Group 1- from two days to one year, Group II- from four to six years of age) undergoing chest PA/AP examinations. At the time of the examination the exposure parameters (kVp, mAs, focal-spot-to-film distance, etc.) and patient information (gender, height, weight and age) were recorded. The radiographic image quality was evaluated by the local radiologist based on the European Guidelines on Quality Criteria for Diagnostic Radiographic Images in Pediatrics. The results showed that the exposure parameters used on newborn patients were in the majority outside the 60-65kV range recommended by the European Guidelines for a good radiographic practice. In the case of examinations of patients with age between 4 and 6 years, 80% were performed with a peak tube voltage within the 60-80 kV range, as recommended by the European Guidelines. The K< values, for both patient groups showed a wide dispersion, ranging from 30.7 uGy to 163.8 uGy for the newborn patients and from 34.5 uGy to 236.6 uGy for infant patients. It is possible to conclude that, in the Brazilian hospitals participating in this project, the radiographic techniques of pediatric chest images and the radiation protection need to be optimized.
机译:这项工作介绍了在巴西进行的儿科胸部检查中所使用的入口表面空气比释动能值(Ke),图像质量和射线照相曝光参数的调查结果。这项研究是国际原子能机构区域项目RLA / 9/057的活动的一部分,该项目的目的是在诊断和介入放射学,核医学和放射治疗中优化对患者的放射防护。这项调查是在巴西的四家医院进行的。研究组包括来自两个年龄组(第1组从两天到一年,第II组从四到六岁)的420名儿科患者,他们接受了胸部PA / AP检查。在检查时,记录了暴露参数(kVp,mAs,焦点到胶片的距离等)和患者信息(性别,身高,体重和年龄)。放射线图像质量由当地放射科医生根据《欧洲儿科诊断放射线图像质量标准指南》进行评估。结果表明,新生儿患者使用的暴露参数多数在欧洲放射线指南推荐的60-65kV范围内,以保证良好的放射学实践。在对4至6岁的患者进行检查的情况下,按照《欧洲指南》的建议,执行80%的峰值管电压在60-80 kV范围内。两组患者的K <值均显示出较宽的离散度,新生儿患者从30.7 uGy到163.8 uGy,婴儿患者从34.5 uGy到236.6 uGy。可以得出结论,在参与该项目的巴西医院中,需要优化儿科胸部图像的放射线照相技术和放射防护。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号