Radiation risk for women in screening mammography examinations in Poland in 2007

机译:2007年波兰筛查钼靶检查中妇女的辐射风险

获取原文

摘要

Doses for women undergoing screening mammography examination should be as low as reasonably achievable, but compatible with the image quality necessary for an adequate diagnosis. The aim of this study was estimation of radiation risk according to "European guidelines for quality assurance in breast cancer screening and diagnosis" [1].Material for this study are the data from 250 screening mammography facilities. For every mammography facility standard average glandular dose for routine exposure were calculated. Furthermore, average glandular doses for individual mammography examination, according to the methods proposed by Dance and others [2], were calculated. Established average glandular doses in 250 mammography facilities range from 0.24 mGy to 15.2 mGy. At only 49 mammography facilities all expositions do not exceed acceptable levels of average glandular dose. At only 25 mammography facilities all expositions do not exceed achievable levels of average glandular dose. Obtained average glandular doses for women undergoing mammography screening examination reach alarming high values and they depend on technical parameters of mammography equipment.
机译:接受乳腺钼靶筛查的妇女的剂量应尽可能低,但应与充分诊断所必需的图像质量相适应。这项研究的目的是根据“欧洲乳腺癌筛查和诊断质量保证指南” [1]估算放射线风险。本研究的材料是来自250台乳腺X线检查设备的数据。对于每个乳腺X线摄影设备,计算出常规暴露的标准平均腺体剂量。此外,根据Dance等人[2]提出的方法,计算了每个乳腺X线摄影检查的平均腺体剂量。在250个乳房X线摄影设备中确定的平均腺体剂量范围为0.24 mGy至15.2 mGy。在仅有的49处乳房X线照相设备上,所有曝光量均未超过可接受的平均腺体剂量水平。在仅有的25个乳房X线照相设备上,所有曝光量均未超过可达到的平均腺体剂量水平。接受乳房X线检查的女性平均腺体剂量达到惊人的高水平,这取决于乳房X线检查设备的技术参数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号