^4fA Laser Interferometer Mercury Manometer used as primary standard for measurement of relative pressures up to 10 kPa

机译:^ 4FA激光干涉仪汞压力计用作测量相对压力的主要标准高达10 KPA

获取原文

摘要

^7fAn improved version, designated MM2, of the MM1 5 kPa full-scale standard manometer of the IMGC-CNR is described. The instrument has been studied and constructed jointly by SPRING Singapore and IMGC-CNR in order to establish a reliable and accurate primary standard for relative gas pressures from 1 Pa to 10 kPa. The new manometer consists of a mercury-filled U-tube (60 mm bore) machined inside a massive stainless-steel body. The measurement of the vertical displacement between the two mercury menisci is made through a commercially heterodyne laser interferometer. Direct reflection of laser beams on mercury menisci is obtained by means of focusing optical devices of the so-called cat's-eye type, as developed at the IMGC (very lightweight floats carrying 1/4" corner cube reflectors are also available for fast but less accurate pressure measurements up to 10kPa). A software was developed by SPRING for instant calculation of pressure value and corrections; using suitable procedures, measurement results of high repeatability have been achieved. The paper presents the pressure uncertainty budget for the new version of manometer. The estimated standard uncertainty spans from 12 mPa at 1 Pa up to 46 mPa at 10 kPa, showing improvement over the prototype in most part of the range. For the purpose of validation, an experimental comparison between MM2 and MM1 has been carried out in 2001-2003. After a description of the adopted transfer standard and of the measurements that were carried out, the comparison results are presented and discussed. The comparison has demonstrated a good agreement and compatibility of measurements between the standard low-range manometers of Singapore and Italy.
机译:^ 7FAN改进的版本,指定MM2,MM1 5 KPA的IMGC-CNR的全尺度标准压力计。通过Spring Singapore和IMGC-CNR共同研究和构建了该仪器,以便为相对气体压力从1Pa到10kPa建立可靠和准确的主要标准。新的压力计包括填充汞的U形管(60 mm孔),在巨大的不锈钢体内加工。通过商业外差激光干涉仪进行两种汞肿瘤之间的垂直位移的测量。通过在IMGC(非常轻质浮子上的Cat'-Eye类型的光学器件上的光学器件通过聚焦光学器件来直接反射MERURY MENISCI的直接反射,如IMGC(携带1/4“角隅棱镜角反射器的速度也可用于快速但更少准确的压力测量高达10kPa)。通过弹簧开发了一种软件,以即时计算压力值和校正;使用了高可重复性的测量结果。本文提出了新版本的压力不确定性预算。估计的标准不确定性在10 kPa的1Pa最高可达46MPa的12MPa中跨越,在大部分范围内显示出原型的改善。为了验证目的,2001年进行了MM2和MM1之间的实验比较-2003。在采用的转移标准和进行的测量的描述之后,提出并讨论了比较结果。比较n已经证明了新加坡和意大利标准低频压力计之间的良好协议和兼容性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号