CT Dose Index Test with Standard Phantoms

机译:CT剂量指数试验标准幻像

获取原文

摘要

The aim of the study was to find out how the measured absorbed dose by using standard dose test method (with CT dose head and body phantoms) depends on the different axial scanning technologies and how they affect the measured value.The results show that using pencil type ionization chamber the measured absorbed dose distribution which determines value of CT dose index (CTDI_(100)) is not equable across the ionization chambers.Uncertainty can reach 30 - 35% of the value during dose measurements of 100 mm long part of the head phantom and up to 50% of the value during measurements of the same length body phantom. It is considered that 5% uncertainty is acceptable during dose measurements in the 40 mm middle range of head phantom and 45 mm in the middle of the range in body phantom.
机译:该研究的目的是了解如何通过使用标准剂量测试方法(CT剂量头和身体幻影)测量的吸收剂量取决于不同的轴向扫描技术以及它们如何影响测量值。结果表明使用铅笔型电离室测量的吸收剂量分布,其确定CT剂量指数(CTDI_(100))的值不等于电离室。unc antamey可以在100mm长部分的剂量测量期间达到30-35%的值。在相同长度体幻影的测量期间,幻影和高达50%的值。考虑到在40mm中间的头部阵体中的剂量测量期间可接受5%的不确定度,在身体幻影的范围内45毫米。