Reliability of densitometry for the assessment of the immediate results after implant of radiolucent coronary artery stents

机译:密度测定法用于评估不透射线冠状动脉支架植入后即刻结果的可靠性

获取原文

摘要

Coronary artery stents have emerged as the most promising device for coronary interventions. The outcome of coronary stenting depends on postprocedural lumen diameter. The aim of the study was to compare two angiographic methods for the quantitation of minimal luminal diameter (MLD) before and after implant of 126 coronary stents. Before angioplasty, the MLD measured by the two techniques correlated (r=0.706) and the mean difference was -0.47/spl plusmn/0.43 mm). After stent implant, the correlation coefficient increased to 0.782 and the mean difference decreases to -0.21/spl plusmn/0.50 mm. Thus both methods are appropriate to measure MLD in coronary segments, but the correlation is higher when the vascular conduit is remodelled by an endovascular prosthesis.
机译:冠状动脉支架已成为最有前途的冠状动脉介入治疗设备。冠状动脉支架置入的结果取决于术后腔直径。该研究的目的是比较两种血管造影方法,以量化在植入126枚冠状动脉支架之前和之后的最小管腔直径(MLD)。血管成形术前,通过两种技术测得的MLD相关(r = 0.706),平均差异为-0.47 / spl plusmn / 0.43 mm。支架植入后,相关系数增加到0.782,平均差减小到-0.21 / spl plusmn / 0.50 mm。因此,这两种方法均适用于测量冠状动脉节段的MLD,但是当血管导管被血管内假体重塑时,相关性更高。