Zero-bias operation and beam energy dependence of diamond sensors

机译:钻石传感器的零偏操作和光束能量依赖性

获取原文
获取原文并翻译 | 示例

摘要

A prototype sensor able to work at zero bias has been fabricated using high quality single crystal diamond and two Schottky junctions. The detector's current in the dark is 3.2±0.2×l0~(-14) A, while the current measured under irradiation with 1.6 Gy·min ~(-1) of 6 and 15 MV x-ray photons is 7.5±0.2×l0~(-10) A. The sensitivity is strongly dependent on the applied voltage whereas at zero bias we have measured 37±2 nOGy~(-1), independent of the beam quality. The sensitivity increases with a sub-linear trend up to 100V, as the sensor photocurrent dependence on voltage. Moreover, the device shows very fast response on the rise and fall transients, with the signal stability under irradiation better than 0.3%. The linearity with the dose rate is spread over three orders of magnitude from 1.3×10~(-3) Gymin~(-1) to 1.2 Gymin~(-1) as measured with Co-60 photons.
机译:使用高质量的单晶金刚石和两个肖特基结制造了一种能够在零偏压下工作的原型传感器。探测器在黑暗中的电流为3.2±0.2×l0〜(-14)A,而在1.6 Gy·min〜(-1)的6和15 MV x射线光子辐照下测得的电流为7.5±0.2×l0 〜(-10)A.灵敏度在很大程度上取决于所施加的电压,而在零偏压下,我们测量到37±2 nOGy〜(-1),与光束质量无关。随着传感器光电流对电压的依赖性,灵敏度随着高达100V的亚线性趋势而增加。此外,该器件对上升和下降瞬态响应非常快,在辐射下的信号稳定性优于0.3%。用Co-60光子测得的剂量率线性关系从1.3×10〜(-3)Gymin〜(-1)到1.2 Gymin〜(-1)分布在三个数量级上。