首页> 外文会议>Instrumentation Science and Technology Vol.2 >Design of Parameters for Relative Devices in Optical Fiber Confocal Scanning Microscopy
【24h】

Design of Parameters for Relative Devices in Optical Fiber Confocal Scanning Microscopy

机译:共焦扫描显微镜中相关设备的参数设计

获取原文

摘要

For the high axial resolution and the completely coherent imaging property despite of the actual size of detector in single mode optical fiber confocal scanning microscopy (FOCSM), the FOCSM had been studied and developed rapidly in recent years. The theoretic scheme and coherent transfer function was described. On the base of coherent transfer function, the relation formula was deduced which indicated the relation between the resolution and the parameters of optical fiber, light wavelength and object lens. In the course of deducing, compared with common microscopy, the result that the FOCSM has higher both plane and axial resolutions was explained. In actual application, the relation formula can instruct the design and choosing of parameters of relative devices (such as optical fiber, light wavelength and object lens) according to index of plane and axial resolution.
机译:尽管在单模光纤共聚焦扫描显微镜(FOCSM)中具有检测器的实际尺寸,但由于具有较高的轴向分辨率和完全相干的成像特性,近年来已对FOCSM进行了研究和快速开发。描述了理论方案和相干传递函数。在相干传递函数的基础上,推导了关系式,该关系式表明了分辨率与光纤,光波长和物镜参数之间的关系。在推论过程中,与普通显微镜相比,解释了FOCSM具有更高的平面和轴向分辨率的结果。在实际应用中,该关系式可以根据平面指数和轴向分辨率指示相关设备(如光纤,光波长和物镜)的设计和参数选择。