An in-house predictive control algorithm has been implemented to control blood glucose level in type I diabetic patients, by controlling the insulin infusion rate to a mechanical pump. The role of the disturbance filter in model-based predictive control is underlined and its possibility to improve control performance is exploited. For comparison purposes, a classic PID controller has been designed via CAD tools. Controller performance was assessed in terms of its ability to track a normoglycemic setpoint (81mg/dL) from initial states of hypo- and hyperglycemia, as well as in response to a meal disturbance. Unconstrained control gave satisfactory results, within imposed constraints. Both strategies are implemented in discrete manner, thus suitable for future implementation in wearable devices.
机译:通过控制对机械泵的胰岛素输注速度,已实施了一种内部预测控制算法来控制I型糖尿病患者的血糖水平。强调了干扰滤波器在基于模型的预测控制中的作用,并探讨了其改善控制性能的可能性。为了比较,已经通过CAD工具设计了经典的PID控制器。根据从低血糖和高血糖症的初始状态以及对进餐障碍的反应中追踪正常血糖设定点(81mg / dL)的能力来评估控制器的性能。在施加的限制内,不受约束的控制给出了令人满意的结果。两种策略均以离散方式实施,因此适合将来在可穿戴设备中实施。