首页> 外文会议>Hard x-ray, gamma-ray, and neutron detector physics XII >Measurements with Coplanar Grid (Cd,Zn)Te Detectors andDevelopment of the GMCA (Gamma-ray analysis digital filter Multi Channel Analyzer)
【24h】

Measurements with Coplanar Grid (Cd,Zn)Te Detectors andDevelopment of the GMCA (Gamma-ray analysis digital filter Multi Channel Analyzer)

机译:使用共面网格(Cd,Zn)Te探测器进行测量和 r n开发GMCA(伽马射线分析数字滤波器多通道分析仪)

获取原文
获取原文并翻译 | 示例

摘要

Gamma-ray spectroscopic systems with high efficiency and energy resolution have become important in many fields. Conventional systems with analog shaping are more and more replaced by digital systems that can provide higher throughput, better energy resolution and better stability. We present the development of our new versatile real time Multi Channel Analyzer with digital signal processing, the GMCA. The device has two independent channels and is compatible with various detectors like Coplanar Grid (Cd,Zn)Te, HPGe, LaBr3, Nal and Csl. Different sensors for temperature, pressure and humidity provide additional information and allow the correction of thermal drifts. We show energy resolution measurements and compare spectra measured with different detectors connected to our system and to commercial systems both analog and digital.
机译:具有高效率和能量分辨率的伽马射线光谱系统在许多领域中已经变得重要。具有模拟整形的常规系统越来越多地被数字系统取代,数字系统可以提供更高的吞吐量,更好的能量分辨率和更好的稳定性。我们介绍了具有数字信号处理功能的新型多功能实时多通道分析仪GMCA的开发。该设备具有两个独立的通道,并兼容各种探测器,例如共面栅(Cd,Zn)Te,HPGe,LaBr3,Nal和Csl。温度,压力和湿度的不同传感器可提供更多信息,并可以校正热漂移。我们展示了能量分辨率测量结果,并比较了使用连接到我们的系统以及模拟和数字系统的不同检测器测得的光谱。