Evaluation of 3D Breast Surface ReconstructionAccuracy Using Non-contact Scanner Images: A Phantom Study

机译:使用非接触式扫描仪图像评估3D乳房表面重建准确性的幻影研究

获取原文
获取原文并翻译 | 示例

摘要

Analysis of the breast surface deformation during mammographic compression will provide insight into breast tissue elastic properties. Our long term goal is to accurately estimate physical properties of breast tissue for the purpose of realistic deformation simulation. As a first step, we analyzed the acquisition accuracy of the reconstructed breast phantom surface using a non-contact laser scanner. The effect of the acquisition angle and the presence of the transparent compression plate covering the phantom were considered. Comparing 48 phantom surfaces, each reconstructed from 4 scanner images acquired at different positions, we observed an average acquisition error of 1.94 mm for images without the compression plate; the average acquisition error for images with the compression plate was 3.26 mm. Performance of two observers showed little variation.
机译:在乳房X线照片压缩期间对乳房表面变形的分析将提供对乳房组织弹性性质的了解。我们的长期目标是准确地估计乳腺组织的物理性质,以实现逼真的变形模拟。第一步,我们使用非接触式激光扫描仪分析了重建的乳房幻像表面的采集精度。考虑了获取角度的影响和覆盖模型的透明压缩板的存在。比较48个幻像表面,每个幻像表面都是从在不同位置获取的4个扫描仪图像重建的,对于没有压缩板的图像,我们观察到平均获取误差为1.94 mm;使用压缩板的图像的平均采集误差为3.26 mm。两名观察员的表现几乎没有差异。