Use of plastic scintillator with fiber optics for detection of radiation with high spatial resolution

机译:使用塑料闪烁体和光纤来检测具有高空间分辨率的辐射

获取原文
获取原文并翻译 | 示例

摘要

The objective fo this paper is to describe a method for detecting x-ray in real time with high spatial resolution. The method uses a 1-meter glass fiber, a 2 mm plastic scintillator, and a collecting device such as photomultiplier tube to respond to the radiation field. This device can be automated for use in any industries, which uses radiation for their processes. The resolution can only be limited by the size of the detector. This plastic scintillator probe has a homogeneous sensitivity and it is water equivalent. The scintillator minimizes dose perturbation in water, and it has a emission peak around 432 nm. The reproducibility of the system has been tested, and the percent variation has been less than 3
机译:本文的目的是描述一种以高空间分辨率实时检测X射线的方法。该方法使用1米长的玻璃纤维,2毫米的塑料闪烁体和诸如光电倍增管的收集装置来响应辐射场。该设备可以自动化用于任何使用辐射进行生产的行业。分辨率只能受检测器尺寸的限制。这种塑料闪烁体探头具有均一的灵敏度,并且相当于水。闪烁体可最大程度地减少水中的剂量扰动,并具有约432 nm的发射峰。系统的可重复性已经过测试,偏差百分比小于3