首页> 外文会议>Biomedical Engineering (ICoBE), 2012 International Conference on >Stability of cementless hole and non-hole hip prosthesis
【24h】

Stability of cementless hole and non-hole hip prosthesis

机译:非骨水泥孔和非孔髋关节假体的稳定性

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Lacking of stability known to be as one of the principal factors contributes to loosening of hip prosthesis. Initial stability of the hip prosthesis related with the magnitude of relative displacement at the femoral bone-prosthesis interface. The present study aimed to investigate the effect hole as additional features on the initial stability. 3D finite element model of femur and prosthesis developed based on CT dataset of a Malaysian patient. Simulations of normal walking condition performed on the models to investigate the relative displacement between the bone and prosthesis interface. The simulations results report that additional hole as a feature on the proximal region of prosthesis, produce highest relative displacement at the proximal region on the medial side.
机译:缺乏稳定性是导致髋关节假体松动的主要因素之一。髋关节假体的初始稳定性与股骨假体界面相对位移的大小有关。本研究旨在调查影响孔作为初始稳定性的附加特征。基于马来西亚患者的CT数据集开发了股骨和假体的3D有限元模型。在模型上进行正常步行条件的仿真,以研究骨骼与假体界面之间的相对位移。模拟结果报告,假体近端区域的附加孔洞在内侧近端区域产生最大相对位移。