首页> 外文会议>Advanced biomedical and clinical diagnostic systems VIII >The VascuLuminator: effectiveness of a near-infrared vessel imaging system as a support in arterial puncture in children
【24h】

The VascuLuminator: effectiveness of a near-infrared vessel imaging system as a support in arterial puncture in children

机译:VascuLuminator:近红外血管成像系统作为支持儿童动脉穿刺的功效

获取原文
获取原文并翻译 | 示例

摘要

A practical near-IR blood vessel imaging system, the 'VascuLuminator', was developed to facilitate the puncturing of blood vessels for different procedures. Technical solutions were found for certain difficulties, such as obtaining a maximum image contrast by reducing the interference of IR light present in the surroundings. In phantom studies it was shown that the device is able to visualize blood vessels of different sizes to a clinically relevant maximum depth. In a preliminary clinical study, the use of the VascuLuminator resulted in decrease of the failure rate in blood withdrawal in young children from 13% to 2% and the laboratory technicians were satisfied with the practical application of the device. After this study, the effectiveness of the VascuLuminator was investigated to facilitate arterial cannulation in a group of children undergoing cardiac surgery. In an ongoing study, 71 children up to 3 years of age were included and time of arterial cannulation, number of punctures and puncture site were recorded. In 38 patients, cannulation was performed without the VascuLuminator and in 33 patients with VascuLuminator by pediatric anesthesiologists. The initial results do not show significant differences in time and in number of punctures with and without the use of the VascuLuminator. However, the VascuLuminator was able to visualize the arteries in most cases. In 11 of the 33 cases, the artery was located by using only the near-infrared image was used, without palpating for a pulse or knowledge of anatomical landmarks. Further clinical studies are needed to identify the patients groups that will benefit the most from VascuLuminator-assisted vessel punctures.
机译:开发了实用的近红外血管成像系统“ VascuLuminator”,以促进针对不同程序的血管穿刺。人们发现了某些困难的技术解决方案,例如通过减少周围环境中红外光的干扰来获得最大的图像对比度。在幻像研究中,该设备能够将不同大小的血管可视化至临床相关的最大深度。在一项初步的临床研究中,使用VascuLuminator可将幼儿抽血的失败率从13%降低到2%,实验室技术人员对该设备的实际应用感到满意。在这项研究之后,对VascuLuminator的有效性进行了研究,以促进一组接受心脏手术的儿童的动脉插管。在一项正在进行的研究中,纳入了71名3岁以下的儿童,并记录了动脉插管时间,穿刺次数和穿刺部位。小儿麻醉医生对38例没有使用VascuLuminator的患者进行插管,对33例使用VascuLuminator的患者进行了插管。在使用和不使用VascuLuminator的情况下,初步结果并未显示出时间和穿刺次数的显着差异。但是,在大多数情况下,VascuLuminator能够可视化动脉。在33例病例中的11例中,仅通过使用近红外图像即可定位动脉,而无需触诊脉搏或了解解剖标志。需要进一步的临床研究来确定将从VascuLuminator辅助血管穿刺中受益最大的患者组。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号