首页> 外文会议>2013 IEEE International Conference on Medical Imaging Physics and Engineering >Detection and evaluation on the accuracy of fiducial tracking with spine auxiliary positioning during IGRT treatment
【24h】

Detection and evaluation on the accuracy of fiducial tracking with spine auxiliary positioning during IGRT treatment

机译:IGRT治疗过程中脊柱辅助定位基准跟踪的准确性检测与评估

获取原文
获取原文并翻译 | 示例

摘要

With spine auxiliary positioning during IGRT treatment, utilize the simulated human phantom and dynamic lung phantom to detect the accuracy of fiducial tracking, and evaluate the value of it. Utilize CT to scan the phantoms which contain films, and design plans for phantoms and spine auxiliary positioning which use 1, 2, 3, 4, 5 and 6 fiducials for tracking, respectively. Each phantom plan shall be repeatedly implemented 3 times. The E2E software is used to analyze the accuracy of irradiation and then the average value is acquired. The results of detecting the simulated human phantom: the accuracy of irradiation with 1 and 2 fiducials for tracking (combining with spine auxiliary positioning) was 1.11 mm and 1.05 mm; with 3, 4, 5 and 6 fiducials, the accuracy was 1.07 mm, 0.92 mm, 0.97 mm and 1.15 mm, respectively. The results of detecting dynamic lung phantom: with 1 and 2 fiducials used for tracking (combining with spine auxiliary positioning), the accuracy of irradiation was 1.03 mm and 0.70 mm; with 3, 4, 5 and 6 fiducials, the accuracy was 0.58 mm, 1.02 mm, 0.65 mm and 0.96 mm, respectively. For the relatively static tumor or moving tumor, if the rotation direction of tumor is consistent with spine, the accuracy of using 1 or 2 fiducials for tracking was the same as using 3 to 6 fiducials. When the number of fiducials is not enough, fiducial tracking along with spine auxiliary positioning can well solve this problem.
机译:在IGRT治疗期间通过脊柱辅助定位,利用模拟的人体模型和动态肺部模型来检测基准跟踪的准确性,并评估其价值。利用CT扫描包含胶片的体模,并设计体模和脊柱辅助定位的设计计划,分别使用1、2、3、4、5和6个基准进行跟踪。每个幻影计划应重复执行3次。使用E2E软件分析辐射的准确性,然后获取平均值。模拟人体模型的检测结果:用1个和2个基准点进行跟踪的跟踪精度(结合脊柱辅助定位)分别为1.11 mm和1.05 mm;使用3个,4个,5个和6个基准时,精度分别为1.07毫米,0.92毫米,0.97毫米和1.15毫米。动态肺部幻象的检测结果:采用1个和2个基准点进行跟踪(结合脊柱辅助定位),照射的准确度分别为1.03 mm和0.70 mm;使用3、4、5和6个基准时,精度分别为0.58 mm,1.02 mm,0.65 mm和0.96 mm。对于相对静止的肿瘤或活动性肿瘤,如果肿瘤的旋转方向与脊柱一致,则使用1个或2个基准进行跟踪的准确性与使用3到6个基准的准确性相同。当基准点数量不够时,基准点跟踪以及脊柱辅助定位可以很好地解决此问题。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号