Test methods and results are presented for the equivalent average input rate of a single-degree-of-freedom gyroscope operated both open loop and with a ternary-logic pulse-torque-to-balance loop during multiaxis angular oscillation. For the open-loop tests, good agreement was obtained with theoretical results. Two-axis testing was performed for oscillations about the Input-Output axes, the Input-Spin axes, and the Spin-Output axes. These tests run in the torque-to-balance mode revealed significant departures from open-loop results in the induced drift rate. An analysis is developed explaining much of the closed-loop data presented. Test data for the gryoscope in a ternary torque-to-balance loop with constant input rates is presented. The tests demonstrate that the instrument rate linearity does not change with interrogation frequency from 3,600 to 14,400 Hz if the torque coil is tuned to offer a resistive load to the current switch. Analysis cited shows that gyroscope lag compensation eliminates multiple pulsing and other equivalent forms of degraded resolution in a wide variety of quantizing loops. This result is test verified for the ternary delta-modulator loop. (Author)