首页> 美国政府科技报告 >Improved Mass-Measurement Accuracy Using a PNB Load Cell Scale
【24h】

Improved Mass-Measurement Accuracy Using a PNB Load Cell Scale

机译:使用pNB称重传感器标度提高质量测量精度

获取原文

摘要

The PNB Load Cell Scale is a Preloaded, Narrow-Band calibration mass comparator. It consists of (1) a frame and servo-mechanism that maintains a preload tension on the load cell until the load, an unknown mass, is sensed, and (2) a null-balance digital instrument that suppresses the cell response associated with the preload, thereby improving the precision and accuracy of the measurements. Ideally, the objects used to set the preload should be replica mass standards that closely approximate the density and mass of the unknowns. The advantages of the PNB scale are an expanded output signal over the range of interest which increases both the sensitivity and resolution, and minimizes the transient effects associated with loading of load cells. An area of immediate and practical application of this technique to nuclear material safeguards is the weighing of UF sub 6 cyliners where in-house mass standards are currently available and where the mass values are typically assigned on the basis of comparison weighings. Several prototypical versions of the PNB scale have been assembled at the US National Bureau of Standards. A description of the instrumentation, principles of measurements, and applications are presented in this paper. (ERA citation 07:013911)