Effect of back scatter radiation on surface dose as measured by TLDs

机译:通过TLD测量的反向散射辐射对表面剂量的影响

获取原文

摘要

A group of scientists at HASYLAB in Germany are doing an Angiography project in parallel with a group of scientists at the Brookhaven National, National Synchrotron Light Source. In order to determine the X-ray dose to a patient, the German group places lithium fluoride thermoluminescent dosimeters (TLD) on the front (upstream) surface of their patient. The found that the TLD dose was higher than the dose calculated from an ionization chamber. An ionization chamber is a device that counts the number of photons when a beam of synchrotron radiation passes through Dr. Wolf-Rainer Dix of HASYLAB found that there is a difference of about 15% between dose data from TLDs put on a water tub and TLDs in the air. The goal of the experiment we did was to understand and maybe match the dose numbers Dix obtained.