A good, functional Hydrogen Fluoride Gasbadge dosimeter has been developed for sampling of airborne HF vapor. The device is small (7.7 cm (times) 5.4 cm (times) 1.9 cm) and can easily and conveniently be worn on one's lapel. It consists of polyethylene and polypropylene parts and a triethanolamine-impregnated polyproylene collection element. It is completely self contained, requiring no pumps, impingers, or sampling tubes. Subsequent to sampling, the collection element is analyzed quickly and easily with a fluoride selective-ion electrode. Laboratory tests were conducted to determine precision, linearity, interference effects, influences of temperature and humidity, and collection element stability over time. Results of the tests indicate that the Abcor Gasbadge HF dosimeter is an excellent passive HF monitor for work spaces, and that results obtained with it are accurate within (plus minus)25%. These results have been corroborated in a field study.