首页> 美国政府科技报告 >Temperature Measurements with Micrometer Spatial Resolution
【24h】

Temperature Measurements with Micrometer Spatial Resolution

机译:微米空间分辨率的温度测量

获取原文

摘要

A novel technique has been developed to measure temperatures with a spatialresolution less than 10 micron. This method uses the temperature-sensitive time decay fluorescence of a phosphor as a surface sensor. The spatial resolution is obtained by using an electron beam to excite individual phosphor grains deposited on the surface of interest. The phosphor selected for its dynamic range and chemical stability is the inorganic compound magnesium fluorogermanate activated with manganese. The time decay constant of its fluorescence decreases monotonically with temperature between -200 to 450 deg C. This method was verified on a cross-sectioned Zener diode. A description of the technique and a temperature map of the sectioned area of an operating diode are presented. Fluorescence, Junction temperature, Semiconductor, Thermometry.