首页> 美国政府科技报告 >Beam-Attenuation Measurement Error Due to Small-Angle Scattering Acceptance
【24h】

Beam-Attenuation Measurement Error Due to Small-Angle Scattering Acceptance

机译:小角度散射验收引起的光束衰减测量误差

获取原文

摘要

Transmissometers always accept a certain portion of the forward scattered light.The error in the beam attenuation coefficient c-lambda due to this acceptance varies with the small angle scattering function, single scattering albedo, and instrumental design. This paper details methods to model error in measurements of c-lambda obtained with cylindrically limited and collimated beam transmissometers. These models are used with real examples of beam attenuation and small angle scattering function data to test how this measurement error varies.