首页> 外文期刊>Physics in medicine and biology. >Reconstruction of a ring applicator using CT imaging: impact of the reconstruction method and applicator orientation
【24h】

Reconstruction of a ring applicator using CT imaging: impact of the reconstruction method and applicator orientation

机译:使用CT成像重建环形涂药器:重建方法和涂药器方向的影响

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The purpose of this study is to investigate whether the method of applicator reconstruction and/or the applicator orientation influence the dose calculation to points around the applicator for brachytherapy of cervical cancer with CT-based treatment planning. A phantom, containing a fixed ring applicator set and six lead pellets representing dose points, was used. The phantom was CT scanned with the ring applicator at four different angles related to the image plane. In each scan the applicator was reconstructed by three methods: ( 1) direct reconstruction in each image (DR), ( 2) reconstruction in multiplanar reconstructed images (MPR) and ( 3) library plans, using pre-defined applicator geometry ( LIB). The doses to the lead pellets were calculated. The relative standard deviation (SD) for all reconstruction methods was less than 3.7% in the dose points. The relative SD for the LIB method was significantly lower ( p < 0.05) than for the DR and MPR methods for all but two points. All applicator orientations had similar dose calculation reproducibility. Using library plans for applicator reconstruction gives the most reproducible dose calculation. However, with restrictive guidelines for applicator reconstruction the uncertainties for all methods are low compared to other factors influencing the accuracy of brachytherapy.
机译:这项研究的目的是调查使用基于CT的治疗计划进行宫颈癌近距离治疗的涂药器重建方法和/或涂药器方向是否影响剂量计算到涂药器周围的点。使用了包含固定环涂抹器套件和代表剂量点的六个铅丸的体模。用环形涂药器以与图像平面相关的四个不同角度对体模进行CT扫描。在每次扫描中,都通过三种方法来重建施加器:(1)在每个图像中直接重建(DR),(2)在多平面重建图像(MPR)中进行重建,以及(3)使用预定义的施加器几何形状(LIB) 。计算铅粒的剂量。所有重建方法的相对标准偏差(SD)在剂量点均小于3.7%。除了两点外,LIB方法的相对SD显着低于DR和MPR方法(p <0.05)。所有施药方向均具有相似的剂量计算重现性。使用图书馆计划进行涂药器重建可提供最可重复的剂量计算。然而,与限制近距离放疗准确性的其他因素相比,在有限制的涂药器重建指导原则下,所有方法的不确定性均较低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号