Three-dimensional dose verification of the clinical application of gamma knife stereotactic radiosurgery using polymer gel and MRI

机译:使用聚合物凝胶和MRI的伽马刀立体定向放射外科手术临床应用的三维剂量验证

获取原文
获取原文并翻译 | 示例
           

摘要

This work seeks to verify multi-shot clinical applications of stereotactic radiosurgery with a Leksell Gamma Knife model C unit employing a polymer gel-MRI based experimental procedure, which has already been shown to be capable of verifying the precision and accuracy of dose delivery in single-shot gamma knife applications. The treatment plan studied in the present work resembles a clinical treatment case of pituitary adenoma using four 8 mm and one 14 mm collimator helmet shots to deliver a prescription dose of 15 Gy to the 50% isodose line (30 Gy maximum dose). For the experimental dose verification of the treatment plan, the same criteria as those used in the clinical treatment planning evaluation were employed. These included comparison of measured and GammaPlan calculated data, in terms of percentage isodose contours on axial, coronal and sagittal planes, as well as 3D plan evaluation criteria such as dose-volume histograms for the target volume, target coverage and conformity indices. Measured percentage isodose contours compared favourably with calculated ones despite individual point fluctuations at low dose contours (e.g., 20%) mainly due to the effect of T2 measurement uncertainty on dose resolution. Dose-volume histogram data were also found in a good agreement while the experimental results for the percentage target coverage and conformity index were 94% and 1.17 relative to corresponding GammaPlan calculations of 96% and 1.12, respectively. Overall, polymer gel results verified the planned dose distribution within experimental uncertainties and uncertainty related to the digitization process of selected GammaPlan output data.
机译:这项工作旨在验证Leksell Gamma刀C型单元采用基于聚合物凝胶MRI的实验程序的立体定向放射外科手术的多镜头临床应用,该实验已被证明能够验证单一剂量给药的精度和准确性。射伽马刀应用。在本研究中研究的治疗计划类似于垂体腺瘤的临床治疗案例,使用四个8 mm和一个14 mm的准直器头盔镜头向15%的等剂量线(最大剂量为30 Gy)提供15 Gy的处方剂量。为了对治疗计划进行实验剂量验证,采用了与临床治疗计划评估中所用的相同标准。这些措施包括根据轴向,冠状和矢状平面上的等剂量轮廓百分比对测量数据和GammaPlan计算数据进行比较,以及3D计划评估标准,例如目标体积,目标覆盖率和合格指数的剂量-体积直方图。尽管在低剂量轮廓(例如20%)处存在单个点波动,但测得的等剂量线轮廓百分比与计算的等剂量线轮廓相比还是令人满意的,这主要是由于T2测量不确定性对剂量分辨率的影响。剂量-体积直方图数据也吻合得很好,而相对于相应的GammaPlan计算结果分别为96%和1.12,目标覆盖率和合格指数的实验结果分别为94%和1.17。总体而言,聚合物凝胶结果验证了实验不确定性和与所选GammaPlan输出数据的数字化过程相关的不确定性内的计划剂量分布。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号