Development of a real-time monitoring system for intra-fractional motion in intracranial treatment using pressure sensors

机译:使用压力传感器开发颅内治疗中的分数运动的实时监控系统

获取原文
获取原文并翻译 | 示例
           

摘要

This study developed a dedicated real-time monitoring system to detect intra-fractional head motion in intracranial radiotherapy using pressure sensors. The dedicated real-time monitoring system consists of pressure sensors with a thickness of 0.6 mm and a radius of 9.1 mm, a thermoplastic mask, a vacuum pillow, and a baseplate. The four sensors were positioned at superior-inferior and right-left sides under the occipital area. The sampling rate of pressure sensors was set to 5 Hz. First, we confirmed that the relationship between the force and the displacement of the vacuum pillow follows Hook's law. Next, the spring constant for the vacuum pillow was determined from the relationship between the force given to the vacuum pillow and the displacement of the head, detected by Cyberknife target locating system (TLS) acquisitions in clinical application. Finally, the accuracy of our system was evaluated by using the 2 x 2 confusion matrix. The regression lines between the force, y, and the displacement, x, of the vacuum pillow were given by y = 3.8x, y = 4.4x, and y = 5.0x when the degree of inner pressure was -12 kPa,-20 kPa, and -27 kPa, respectively. The spring constant of the vacuum pillow was 1.6 N mm(-1) from the 6D positioning data of a total of 2999 TLS acquisitions in 19 patients. Head motions of 1 mm, 1.5 mm, and 2 mm were detected in real-time with the accuracies of 67%, 84%, and 89%, respectively. Our system can detect displacement of the head continuously during every interval of TLS with a resolution of 1-2 mm without any radiation exposure.
机译:这项研究开发了一种专用的实时监测系统,以使用压力传感器检测颅内放疗中的颅内运动。专用的实时监控系统由厚度为0.6毫米,半径为9.1毫米的压力传感器,热塑性面罩,真空枕和底板组成。四个传感器分别位于枕骨区域下方的上下左右。压力传感器的采样率设置为5 Hz。首先,我们确认真空枕头的力和位移之间的关系遵循胡克定律。接下来,根据在临床应用中通过射波刀目标定位系统(TLS)采集的数据,通过施加在真空枕上的力与头部的位移之间的关系来确定真空枕的弹簧常数。最后,我们使用2 x 2混淆矩阵评估了系统的准确性。当内部压力为-12 kPa,-20时,真空枕的力y和位移x之间的回归线为y = 3.8x,y = 4.4x和y = 5.0x。 kPa和-27 kPa。根据19个患者总共2999个TLS采集的6D定位数据,真空枕的弹簧常数为1.6 N mm(-1)。实时检测到1 mm,1.5 mm和2 mm的头部运动,其准确度分别为67%,84%和89%。我们的系统可以在TLS的每个间隔内连续检测到头部的位移,分辨率为1-2 mm,而没有任何辐射。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号