Monte Carlo dose calculations using MCNP4C and EGSnrc/BEAMnrc codes to study the energy dependence of the radiochromic film response to beta-emitting sources

机译:使用MCNP4C和EGSnrc / BEAMnrc代码进行蒙特卡洛剂量计算,以研究放射变色膜对β发射源的能量依赖性

获取原文
获取原文并翻译 | 示例
           

摘要

The energy dependence of the radiochromic film (RCF) response to beta-emitting sources was studied by dose theoretical calculations, employing the MCNP4C and EGSnrc/BEAMnrc Monte Carlo codes. Irradiations with virtual monochromatic electron sources, electron and photon clinical beams, a P-32 intravascular brachytherapy (IVB) source and other beta-emitting radioisotopes (Re-188, Y-90, Sr-90/Y-90, P-32) were simulated. The MD-55-2 and HS radiochromic films (RCFs) were considered, in a planar or cylindrical irradiation geometry, with water or polystyrene as the surrounding medium. For virtual monochromatic sources, a monotonic decrease with energy of the dose absorbed to the film, with respect to that absorbed to the surrounding medium, was evidenced. Considering the IVB P-32 source and the MD-55-2 in a cylindrical geometry, the calibration with a 6 MeV electron beam would yield dose underestimations from 14 to 23%, increasing the source-to-film radial distance from 1 to 6 mm. For the planar beta-emitting sources in water, calibrations with photon or electron clinical beams would yield dose underestimations between 5 and 12%. Calibrating the RCF with Sr-90/ 90Y, the MD-55-2 would yield dose underestimations between 3 and 5% for P-32 and discrepancies within 2% for 188Re and 90Y, whereas for the HS the dose underestimation would reach 4% with 188Re and 6% with P-32.
机译:通过使用MCNP4C和EGSnrc / BEAMnrc蒙特卡罗代码进行剂量理论计算,研究了放射性变色膜(RCF)对β发射源的能量依赖性。用虚拟单色电子源,电子和光子临床束,P-32血管内近距离放射治疗(IVB)以及其他发射β射线的放射性同位素(Re-188,Y-90,Sr-90 / Y-90,P-32)照射被模拟。认为MD-55-2和HS辐射致变色膜(RCF)以平面或圆柱状的辐射几何形状,以水或聚苯乙烯为周围介质。对于虚拟单色源,相对于周围介质吸收的能量,膜吸收的能量随能量的增加单调减少。考虑到IVB P-32辐射源和MD-55-2呈圆柱形,使用6 MeV电子束进行校准会导致剂量低估从14%到23%,从而使源到膜的径向距离从1增长到6毫米对于水中的平面β发射源,用光子或电子临床光束进行校准会导致剂量低估5%至12%。用Sr-90 / 90Y校准RCF,对于P-32,MD-55-2的剂量低估3%至5%,对于188Re和90Y的差异2%以内,而HS的剂量低估将达到4%使用188Re和6%使用P-32。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号