A model to calculate the induced dose rate around an 18 MV ELEKTA linear accelerator

机译:用于计算18 MV ELEKTA线性加速器周围的诱导剂量率的模型

获取原文
获取原文并翻译 | 示例
           

摘要

The dose rate due to activity induced by (γ, n) reactions around an ELEKTA Precise accelerator running at 18 MV is reported. A model to calculate the induced dose rate for a variety of working practices has been derived and compared to the measured values. From this model, the dose received by the staff using the machine can be estimated. From measured dose rates at the face of the linear accelerator for a 10 * 10 cm~2 jaw setting at 18 MV an activation coefficient per MU was derived for each of the major activation products. The relative dose rates at points around the linac head, for different energy and jaw settings, were measured. Dose rates adjacent to the patient support system and portal imager were also measured. A model to calculate the dose rate at these points was derived, and compared to those measured over a typical working week. The model was then used to estimate the maximum dose to therapists for the current working schedule on this machine. Calculated dose rates at the linac face agreed to within ± 12% of those measured over a week, with a typical dose rate of 4.5μSv h~(-1) 2 min after the beam has stopped. The estimated maximum annual whole body dose for a treatment therapist, with the machine treating at only 18MV, for 60 000 MUs per week was 2.5mSv. This compares well with value of 2.9mSv published for a Clinac 21EX. A model has been derived to calculate the dose from the four dominant activation products of an ELEKTA Precise 18MV linear accelerator. This model is a useful tool to calculate the induced dose rate around the treatment head. The model can be used to estimate the dose to the staff for typical working patterns.
机译:据报道由于在18 MV下运行的ELEKTA Precise加速器周围的(γ,n)反应诱导的活性导致的剂量率。已经得出了用于计算各种工作实践的诱导剂量率的模型,并将其与测量值进行了比较。根据该模型,可以估计工作人员使用机器的剂量。从在18 MV下10 * 10 cm〜2的颌骨设置的直线加速器表面的测量剂量率,得出每个主要活化产物的每MU活化系数。测量了针对不同能量和颌骨设置的直线加速器头部周围各点的相对剂量率。还测量了患者支持系统和门禁成像仪附近的剂量率。得出了在这些点上计算剂量率的模型,并将其与典型工作周中测得的模型进行比较。然后,使用该模型估算该机器当前工作时间表中治疗师的最大剂量。直线加速器面部的计算剂量率同意在一周内测量的剂量率的±12%以内,典型的剂量率在光束停止后2分钟为4.5μSvh〜(-1)。对于仅使用18MV的机器进行治疗的治疗师,每周估计60,000 MU的年度最大全身剂量为2.5mSv。这与Clinac 21EX的2.9mSv值相比较。已经从ELEKTA Precise 18MV线性加速器的四个主​​要激活产物中得出了一个模型来计算剂量。该模型是计算治疗头周围的诱导剂量率的有用工具。该模型可用于估计典型工作模式下员工的剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号