首页> 外文期刊>Sleep medicine >Pressure stability with CPAP devices: A bench evaluation.
【24h】

Pressure stability with CPAP devices: A bench evaluation.

机译:CPAP设备的压力稳定性:基准评估。

获取原文
获取原文并翻译 | 示例
       

摘要

BACKGROUND: Continuous positive airway pressure (CPAP) maintains a constant pressure to reduce the patient's work of breathing (WOB). The aim of this study was to measure the additional WOB imposed by four current CPAP devices during simulation of a difficult but commonly encountered clinical situation. METHOD: Flow contour, respiratory system compliance and total lung-airway resistance of a patient under CPAP were simulated. The devices were tested at a CPAP of 15cm H(2)O with a heated humidifier and a nasal pillow, which increased circuitry resistance and with and without a simulated unintentional leak. RESULTS: With no leak, positive end-expiratory pressure (PEEP) at the interface varied across devices from 14.0 to 15.3cm H(2)O. With a leak of 1L/s, PEEP varied from 11.5 to 17.1cm H(2)O. Imposed inspiratory WOB ranged from less than 0.1J/min to 0.45J/min with no leak, and the range broadened with leaking. Findings were similar for the imposed expiratory WOB. CONCLUSION: The performances of CPAP devices are variable. The device that calibrated for the pressure loss in the circuitry under dynamic conditions and made appropriate pressure adjustments outperformed the other devices.
机译:背景:持续的气道正压(CPAP)保持恒定的压力,以减少患者的呼吸功(WOB)。这项研究的目的是在模拟困难但经常遇到的临床情况时,测量由四个当前CPAP设备施加的额外WOB。方法:模拟CPAP患者的血流轮廓,呼吸系统顺应性和总肺气道阻力。在15cm H(2)O的CPAP上使用加热的加湿器和鼻枕对设备进行了测试,这增加了电路电阻,并且有或没有模拟的意外泄漏。结果:在没有泄漏的情况下,接口处的呼气末正压(PEEP)在不同设备之间从14.0到15.3cm H(2)O不等。泄漏速度为1L / s,PEEP从11.5到17.1cm H(2)O不等。施加的吸气WOB范围从小于0.1J / min到0.45J / min,无泄漏,并且随着泄漏而扩大了范围。强制呼气WOB的发现相似。结论:CPAP设备的性能是可变的。在动态条件下对电路中的压力损失进行校准并进行适当压力调节的设备优于其他设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号