Aim: Aim of this retrospective study was to analyze the effect of closed-loop stimulation (CLS) and DDD pacing mode on autonomic balance, and to evaluate heart rate variability (HRV) during CLS stimulation. Methods and Materials: Autonomic balance was estimated by a 24-hour HRV analysis of paced and spontaneous beats in patients implanted with a dual-chamber pacemaker (Inos2+ CLS - Biotronik GmbH, Berlin, Germany) and randomly assigned to CLS or DDD pacing mode. Patients underwent two 24-hour electrocardiogram Holter recordings at the end of each 3-month pacing mode period. Each Holter recording was automatically scanned to extract sequences of consecutive beats of the same type [atrial paced (Ap)-sequence and atrial spontaneous (As)-sequence], lasting at least 130 beats. Results: Eight hundred and ten sequences were extracted from 15 patients, and the following spectral parameters were evaluated during both CLS and DDD mode: the total power (variance), the absolute and percentage (relative to the total power) powers of the low frequency (LF, 0.04-0.15 Hz) and high frequency (HF, 0.15-0.4 Hz) components, as well as the LF/HF power ratio. Discussion: The two main findings of this study were: in all the patients, CLS seems to mimic short-term physiological HRV, although the variability (total power) was lower than that relative to the spontaneous beats; the HRV of the spontaneous beats had an higher LF/HF when the pacemaker was programmed as DDD respect to CLS, consistent with a shift toward sympathetic predominance.
机译:目的:这项回顾性研究的目的是分析闭环刺激(CLS)和DDD起搏模式对自主神经平衡的影响,并评估CLS刺激期间的心率变异性(HRV)。方法和材料:通过对植入双腔起搏器(Inos2 + CLS-Biotronik GmbH,Berlin,Germany)并随机分配为CLS或DDD起搏模式的患者的起搏和自发搏动进行24小时HRV分析来估计自主平衡。在每个3个月的起搏模式周期结束时,患者接受两次24小时心电图动态心电图记录。自动对每个动态心电记录进行扫描,以提取相同类型的连续搏动序列[心房起搏(Ap)序列和心房自发(As)序列],持续至少130次。结果:从15位患者中提取了810个序列,并在CLS和DDD模式下评估了以下频谱参数:低频的总功率(方差),绝对功率和百分比(相对于总功率) (LF,0.04-0.15 Hz)和高频(HF,0.15-0.4 Hz)组件,以及LF / HF功率比。讨论:这项研究的两个主要发现是:在所有患者中,CLS似乎模仿了短期生理性HRV,尽管变异性(总功率)相对于自发搏动要低。当将起搏器编程为相对于CLS的DDD时,自发搏动的HRV具有较高的LF / HF,这与向交感神经的转变一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号