首页> 外文期刊>PACE: Pacing and clinical electrophysiology >Telemetry guided pacemaker programming: impact of output amplitude and the use of low threshold leads on projected pacemaker longevity.
【24h】

Telemetry guided pacemaker programming: impact of output amplitude and the use of low threshold leads on projected pacemaker longevity.

机译:遥测引导起搏器编程:输出幅度的影响以及使用低阈值导线对起搏器的预期寿命。

获取原文
获取原文并翻译 | 示例
           

摘要

In a prospective study, a low threshold screw-in electrode (Medtronic 5078, group I, n = 9) was compared to a conventional active fixation lead (Biotronik Y60BP, group II, n = 9) to investigate whether lower pacing thresholds really translate into longer projected service life of the pacemaker. The leads were implanted in the atrium and were connected to a dual chamber pacing system which included the same ventricular lead (Medtronic 5024) and the same pulse generator model (Intermedics 294-03) in both groups. Eighteen months after implantation, atrial and ventricular pacing thresholds were measured as the charge delivered per pulse [microC] at 0.5, 1.0, 1.5, 2.0, and 3.5 V, respectively. For chronic output programming in both channels, patients capturing at 0.5 V were set to 1.0 V, those capturing at 1.5 V were permanently programmed to 2.0 V with the double of the charge threshold as the safety margin for pacing ("safety charge"). A combination of atrial and ventricular output settings was optimal, if it resulted in minimum battery current drain (microA] as measured by pacemaker telemetry. In both groups, current consumption [microA] decreased significantly as output amplitude was decreased, exhibiting its lowest value at 1.0 V in either channel. All ventricular leads could be programmed to the optimum output amplitude of 1.0 V in groups 1 and 2. As the 2:1 "safety charge" values were almost identical, the ventricular channel essential contributes the same amount to the battery drain of the pacing system in both groups. In the atrium, all patients of group 1 could be programmed to the optimum output amplitude of 1.0 V with an average pulse duration of 0.42 +/- 0.15 ms. In group 2, however, all patients had to be programmed to 2.0 V with a mean pulse width of 0.52 +/- 0.15 ms. With the atrial and ventricular output being optimized, the average battery drain of the whole pacing system was 12.19 +/- 0.63 microA in group 1 versus 14.42 +/- 0.32 microA in group 2 (P < 0.001). As patients were chronically programmed to these output settings, this difference translates into a clinically relevant gain in projected pacemaker longevity of 17 months or 18.3% (121 +/- 4 vs. 104 +/- 2 months; P < 0.001). Thus, programming a 2:1 safety margin in terms of charge and optimizing the output parameters by real-time telemetry of the battery current is a useful approach to reduce battery current drain. Making the most of modern lead technology with a different performance in only one channel of an otherwise identical DDD pacing system translates into a significant prolongation of projected pacemaker service life which is of great importance with the increasing awareness of health care expenditures. The gain in projected longevity is mainly due to the option of reducing the output amplitude which is still significantly beneficial well below the nominal voltage of the power source.
机译:在一项前瞻性研究中,将低阈值旋入式电极(Medtronic 5078,I组,n = 9)与传统的主动固定引线(Biotronik Y60BP,II组,n = 9)进行了比较,以研究较低的起搏阈值是否真正转化延长起搏器的预期使用寿命。导线被植入心房,并连接到双室起搏系统,该系统包括两组相同的心室导线(Medtronic 5024)和相同的脉冲发生器模型(Intermedics 294-03)。植入后18个月,心房和心室起搏阈值分别以每脉冲[microC]在0.5、1.0、1.5、2.0和3.5 V下输送的电荷进行测量。对于两个通道的长期输出编程,将以0.5 V捕获的患者设置为1.0 V,以1.5 V捕获的患者被永久编程为2.0 V,并且充电阈值的两倍作为起搏的安全裕度(“安全充电”)。如果通过起搏器遥测测得的电池电流消耗最小(microA),则心房和心室输出设置的组合是最佳的;两组中,电流消耗[microA]随输出幅度的降低而显着降低,其最低值出现在在任一通道中均为1.0V。在第1组和第2组中,所有心室引线都可以设置为最佳输出幅度1.0V。由于2:1“安全电荷”值几乎相同,因此心室通道对信号的贡献量相同。两组起搏系统的电池耗电量。在心房中,可以将第1组的所有患者编程为最佳输出幅度为1.0 V,平均脉冲持续时间为0.42 +/- 0.15 ms。患者必须设置为2.0 V,平均脉冲宽度为0.52 +/- 0.15 ms,在优化心房和心室输出的情况下,整个起搏系统的平均电池消耗为12.19 +/- 0.63 microA。第1组与第2组的14.42 +/- 0.32 microA(P <0.001)。由于将患者长期编程为这些输出设置,因此这种差异转化为临床预期的起搏器寿命为17个月或18.3%(121 +/- 4 vs. 104 +/- 2个月; P <0.001)。因此,就充电而言编程2:1的安全裕度并通过电池电流的实时遥测优化输出参数是减少电池电流消耗的有用方法。在其他方面相同的DDD起搏系统的仅一个通道中,充分利用具有不同性能的现代领先技术,可以显着延长预计的起搏器使用寿命,这对提高人们对医疗保健支出的认识日益重要。预计寿命的增加主要是由于降低输出幅度的选择,这仍然大大低于电源的标称电压。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号