...
首页> 外文期刊>PACE: Pacing and clinical electrophysiology >Extensive variation in the signal amplitude of the atrial floating VDD pacing electrode.
【24h】

Extensive variation in the signal amplitude of the atrial floating VDD pacing electrode.

机译:心房浮动VDD起搏电极的信号幅度发生广泛变化。

获取原文
获取原文并翻译 | 示例
           

摘要

The dependence of atrial signal amplitude on the site of the atrial sensing dipole of a single-pass lead was examined in 29 patients. The vertical location of the dipole was documented in supine fluoroscopy during quiescent and deep breathing and in upright chest roentgenogram, and was expressed as a proportion of the right atrial height. As the group average, the atrial signal amplitude was equal when tested in supine, sitting, standing, and right- or left-side positions in follow-up determinations. The signal amplitude varied markedly between postures, showing a coefficient of variation of 45% +/- 24% within the group. Coefficient of variation within the 6-month follow-up period in each tested position ranged from 31%-44%. Correlation between postures was weak (range of r = 0.53-0.81). Vertical location of the atrial dipole had no relationship to the signal amplitude. At least in one posture or test occasion the atrial signal amplitude was very low, < or = 0.35 mV in 20 patients, and below detection limit (0.25 mV) in 5 patients. Programmed to high sensitivity, atrial undersensing was rare in ambulatory electrocardiography, ranging from 0-9,000 missed atrial beats (0%-8%), with a median of 100 beats/24 hours. In conclusion, temporary variation in atrial signal amplitude is extensive. Despite occasionally measured large signal amplitudes atrial sensitivity in single lead VDD pacemakers should be programmed high, and if poor atrial tracking is suspected, other methods besides routine sensitivity testing should be adapted.
机译:在29例患者中检查了心房信号幅度对单次通过导线的心房感应偶极部位的依赖性。静止和深呼吸时仰卧位透视检查以及胸部X线检查均记录了偶极子的垂直位置,并表示为右心房高度的一部分。作为组的平均值,在后续检查中以仰卧位,坐位,站立位以及右侧或左侧位置测试时,心房信号幅度相等。信号幅度在姿势之间显着变化,显示组内的变化系数为45%+/- 24%。每个测试位置的6个月随访期内的变异系数范围为31%-44%。姿势之间的相关性较弱(r = 0.53-0.81的范围)。心房偶极子的垂直位置与信号幅度无关。至少在一种姿势或测试情况下,心房信号幅度非常低,在20位患者中≤0.35 mV,在5位患者中低于检测极限(0.25 mV)。程序设置为高灵敏度,在动态心电图检查中很少发生心房感觉减退,心律失常为0-9,000次(0%-8%),中位数为100次/ 24小时。总之,心房信号幅度的暂时变化是广泛的。尽管偶尔会在单引线VDD起搏器中测得较大的信号幅度,但心房敏感性应设置为较高,如果怀疑房颤跟踪不良,则应采用除常规敏感性测试以外的其他方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号