The effect of sample temperature on the emission line intensification mechanisms in orthogonal double-pulse Laser Induced Breakdown Spectroscopy

机译:正交双脉冲激光诱导击穿光谱中样品温度对发射线增强机理的影响

获取原文
获取原文并翻译 | 示例

摘要

Orthogonal double-pulse laser ablation on aluminum-based alloy target at atmospheric pressure was investigated by time-resolved optical emission spectroscopy. Studies were performed as a function of the distance from the first pulse plasma induced in air to the sample surface. Double-pulse experimental results were compared to single-pulse laser ablation experiments carried out in a heated target to up to 500 °C. Emission line intensification mechanisms are discussed in terms of the sample heating by the pre-ablation pulse. Crater volume was also determined via confocal microscopy for both, double pulse and target heating, experiments. It was found that the emission intensification is mainly due to an enhancement of the ablated mass. The mechanisms for the mass removal increasing are discussed.
机译:通过时间分辨光发射光谱研究了大气压下铝基合金靶材上的正交双脉冲激光烧蚀。根据从空气中感应出的第一脉冲等离子体到样品表面的距离进行研究。将双脉冲实验结果与在高达500°C的加热靶材上进行的单脉冲激光烧蚀实验进行了比较。就通过预烧蚀脉冲加热样品的方式讨论了发射线增强机制。还通过共聚焦显微镜确定了双脉冲和靶加热实验的弹坑体积。发现发射增强主要是由于烧蚀质量的增加。讨论了去除质量增加的机理。