首页> 外文期刊>Spectrochimica Acta, Part B. Atomic Spectroscopy >Determination of arsenic in gold by flow injection inductively coupled plasma mass spectrometry with matrix removal by reductive precipitation
【24h】

Determination of arsenic in gold by flow injection inductively coupled plasma mass spectrometry with matrix removal by reductive precipitation

机译:流动注射电感耦合等离子体质谱-还原沉淀去除基体法测定金中的砷。

获取原文
获取原文并翻译 | 示例

摘要

Arsenic was determined in gold by flow injection hydride generation inductively coupled plasma-mass spectrometry following a batch mode reductive precipitation removal of the interfering gold matrix. A solution of potassium iodide, L-ascorbic acid, and hydrochloric acid was used as the reluctant. The recovery of gold by precipitation and filtration was 99 ± 3%. The detection limit for arsenic in gold was 55 ng g~(-1) in the solid. The concentration of arsenic that was determined in the Royal Canadian Mint gold sample FAU-10 was 29.7 μg g~(-1) in the solid; this value was indistinguishable, with 95% confidence, from values determined at the Royal Canadian Mint by graphite furnace atomic absorption spectrometry and by inductively coupled plasma-massspectrometry. The standard deviation for four replicate determinations of the arsenic in FAU-10 was 0.972 μg g~(-1) in the solid.
机译:干扰金基质的间歇模式还原沉淀去除后,通过流动注射氢化物发生电感耦合等离子体质谱法测定金中的砷。碘化钾,L-抗坏血酸和盐酸的溶液用作该磁阻剂。通过沉淀和过滤回收的金为99±3%。固体中金的砷的检出限为55 ng g〜(-1)。加拿大皇家铸币厂黄金样品FAU-10中测定的砷浓度为29.7μgg〜(-1);该值与加拿大皇家铸币局通过石墨炉原子吸收光谱法和电感耦合等离子体质谱法测定的值具有95%的置信度相近。固体中FAU-10中四次重复测定砷的标准偏差为0.972μgg〜(-1)。