首页> 外文期刊>Spectrochimica Acta, Part B. Atomic Spectroscopy >More efficient ion transport for inductively coupled plasma mass spectrometry
【24h】

More efficient ion transport for inductively coupled plasma mass spectrometry

机译:用于电感耦合等离子体质谱的更高效的离子传输

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A modification of the ion-lens system in an inductively coupled plasma mass spectrometer has been studied in an attempt to improve the ion transport efficiency. The modification consists of the placement of a metal rod down the axis of the lens-stack.The rod is suspended in place via a ceramic insulator inserted through the photon stop A negative potential is applied to the rod to help focus the positively charged ions into the quadrupole Pnor to the implementation of this modification, a simple model was developed to determine if the voltages required would be practical. The resulting simulated ion trajectories showed that focusing occurs with applied voltages on the order of -10 V, which was approximately the optimal voltage found experimentallyThe experimental data presented show that, with the modification, the limits of detection for 15 elements (m/z ranging from 24 through 238) improved by factors ranging from two to ten times over those of the unmodified instrument.
机译:为了提高离子传输效率,已经研究了在电感耦合等离子体质谱仪中对离子透镜系统的改进。修改包括将金属棒放置在镜头堆栈的轴上。通过插入光子挡块的陶瓷绝缘子将棒悬挂在适当位置。将负电势施加到棒上,以帮助将带正电的离子聚焦到为了实现此修改的四极杆Pnor,开发了一个简单的模型来确定所需的电压是否可行。所得的模拟离子轨迹表明,聚焦发生在施加电压约为-10 V的情况下,这大约是实验中发现的最佳电压。实验数据显示,经过修改,可以检测15个元素的检测极限(m / z范围(从24到238)比未修改的仪器提高了2到10倍。