Teflon Might Be a Factor Accounting for a Failed Microvascular Decompression in Hemifacial Spasm: A Technical Note

机译:铁氟龙可能是导致面肌痉挛微血管减压失败的一个因素:技术说明

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Although Teflon is widely adopted for micro vascular decompression (MVD) surgery, it has never been addressed for failure analysis. This study analyzed the reasons for failed MVDs with emphasis on the Teflon sponge. Methods: Among the 685 hemifacial spasm cases between 2010 and 2014, 31 were reoperated on within a week because of unsatisfactory outcome, which was focused on in this study. Intraoperative findings regarding Teflon inserts of these repeat MVDs were reviewed. Results: Among the 38 without satisfactory outcomes, 31 underwent repeat MVDs, and they were all spasm free afterwards. Eventually, the final cure rate was 99.2%. It was found in the repeat MVDs that the failure was attributable to the Teflon insert in most of the cases (74.2%) directly or indirectly. It was caused by improper placement (47.8%), inappropriate size (34.8%) and unsuitable shape (17.4%) of the Teflon sponge. Conclusion: Although it is not difficult for an experienced neurosurgeon to discover a neurovascular conflict during the MVD process, the size, shape and location of the Teflon sponge should not be ignored. Basically, the Teflon insert is used to keep the offending artery away from the facial nerve root rather than to isolate it. Therefore, the ideal Teflon sponge should be just small enough to produce a neurovascular separation. (C) 2016 S. Karger AG, Basel
机译:背景:尽管特氟龙广泛用于微血管减压(MVD)手术,但从未用于失败分析。这项研究分析了MVD失败的原因,重点是聚四氟乙烯海绵。方法:在2010年至2014年间的685例面肌痉挛病例中,有31例因转归不理想而在一周内再次手术,本研究重点探讨这种情况。回顾了有关这些重复MVD的聚四氟乙烯插入物的术中发现。结果:在38例结果不理想的患者中,有31例接受了重复MVD,之后均无痉挛。最终,最终治愈率为99.2%。在重复的MVD中,在大多数情况下直接或间接地将故障归因于Teflon插件(74.2%)。这是由于特氟隆海绵放置不当(47.8%),尺寸不合适(34.8%)和形状不合适(17.4%)引起的。结论:尽管有经验的神经外科医生在MVD过程中发现神经血管冲突并不困难,但不应忽略Teflon海绵的尺寸,形状和位置。基本上,Teflon插入物用于使病变动脉远离面神经根,而不是将其隔离。因此,理想的聚四氟乙烯海绵应足够小以产生神经血管分离。 (C)2016 S.Karger AG,巴塞尔

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号