Comparison of methods for determining the presence and extent of anterior lumbar interbody fusion.

机译:比较确定前腰椎椎间融合的存在和程度的方法。

获取原文
获取原文并翻译 | 示例
           

摘要

STUDY DESIGN: Titanium alloy interbody fusion devices with autogenous bone were placed in the L5-L6 disc space of 31 adult pig-tailed monkeys through an anterolateral (retroperitoneal) approach. Anteroposterior and lateral radiographs, CT imaging, and histologic analysis of the specimens were performed. OBJECTIVES: This study compared the accuracy of plain film radiographs and CT imaging for determining bony fusion of a titanium interbody device implanted in a non-human primate model. The accuracy of the assessments was determined by comparison to histologic analysis. SUMMARY OF BACKGROUND DATA: Interbody fusion assessment is often difficult to compare in clinical studies because of differences in definition of fusion criteria. In addition, the accuracy of plain film radiographs and CT imaging assessments of fusion are debated because of device material radiopacity and introduction of artifacts. METHODS: A uniform grading system evaluating both the presence and extent of bony fusion was applied to all evaluation techniques. Matched-pair nonparametric t tests were used to determine differences in scoring. RESULTS: The radiographic and histologic presence of fusion grades was equivalent in only 13 of 29 cases (45%), while the CT imaging was equivalent to histologic assessment in 24 of 29 cases (83%). However, the extent of bony fusion in CT imaging and histologic assessment was equivalent in only 4 of 29 cases (14%). Grading of CT images significantly overestimated the extent of fusion. CONCLUSIONS: This study demonstrated CT imaging techniques to be superior to plain film radiographs in determining the presence of bony fusion. However, CT imaging did not accurately determine the extent of bony fusion present as confirmed by histologic analysis.
机译:研究设计:通过前外侧(腹膜后)方法将具有自体骨的钛合金椎间融合器放置在31只成年猪尾猴的L5-L6椎间盘间隙中。进行前后位和侧位X线照片,CT成像以及标本的组织学分析。目的:本研究比较了普通胶片X射线照片和CT成像在确定植入非人类灵长类动物模型中的钛椎间融合器的骨融合时的准确性。通过与组织学分析比较来确定评估的准确性。背景数据摘要:由于融合标准的定义不同,体内融合评估通常在临床研究中难以比较。此外,由于装置材料的射线不透性和伪影的引入,对平片X射线照片的准确性和融合的CT成像评估提出了争议。方法:对骨融合的存在和程度进行评估的统一评分系统应用于所有评估技术。配对非参数t检验用于确定评分差异。结果:29例中只有13例(45%)具有融合级别的影像学和组织学表现,而29例中有24例(83%)相当于CT的组织学评估。然而,在29例中只有4例(14%)在CT成像和组织学评估中达到了骨融合的程度。 CT图像的分级显着高估了融合程度。结论:这项研究表明,CT成像技术在确定骨融合的存在方面优于平片。然而,如通过组织学分析所证实的,CT成像不能准确确定存在的骨融合程度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号