Cause for inadequate clearance: hemodialysis catheter migration into the coronary sinus.

机译:清除不充分的原因:血液透析导管移入冠状窦。

获取原文
获取原文并翻译 | 示例
           

摘要

Complications associated with central dialysis catheters prove to be an important source of morbidity that challenge patients and clinicians alike. While thrombosis, stenosis, and infection remain the most common threats to functioning central dialysis catheters, malposition of such devices may similarly result in serious consequences. Despite advanced techniques used to confirm initial catheter placement, the very nature of these catheters entails frequent manipulation for vascular access and therefore they may migrate after repeated use. When compared with the loss of patency and infection, complications involving malposition or migration may present in a more insidious fashion, such as a gradual decline in achievable blood flow rate or observed clearance over time. We describe a patient who acutely developed both clinical and biochemical evidence of inadequate clearance after several months of suboptimal catheter function. When the central dialysis catheter was examined fluoroscopically, it was found to have migrated into the coronary sinus.
机译:事实证明,与中央透析导管相关的并发症是发病率的重要来源,对患者和临床医生均构成挑战。尽管血栓形成,狭窄和感染仍然是对中央透析导管起作用的最常见威胁,但此类设备的位置不当可能同样会导致严重后果。尽管采用了先进的技术来确定导管的初始位置,但这些导管的本质还是需要对血管通路进行频繁操作,因此,在重复使用后它们可能会迁移。与通畅性和感染的丧失相比,涉及错位或迁移的并发症可能以更隐蔽的方式出现,例如可达到的血液流速逐渐下降或随着时间的推移观察到的清除率。我们描述了一个患者,在几个月的最佳导管功能后,其急性清除率不足的临床和生化证据迅速发展。透视检查中央透析导管时,发现它已迁移到冠状窦中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号