Protrusio of a Ceramic Femoral Head Through the Acetabular Metallic Shell, Extensive Metallosis and 'Bubble Sign'

机译:通过髋臼金属外壳,广泛的金属化和“气泡征”的陶瓷股骨头突出

获取原文
获取原文并翻译 | 示例

摘要

A 24-year-old patient with a history of juvenile rheumatoid arthritis underwent a primary cementless left total hip arthroplasty (THA). The original THA consisted of an Optifix 54 cup with a 3-mm thick polyethylene liner, an Optifix size 4 stem (Smith & Nephew Richards, Memphis, Tennessee), and a Biolox aluminum 32-mm femoral head. Fourteen years later, radiographs demonstrated extensive wear of the polyethylene liner resulting in direct articulation and abrasion wear of the ceramic femoral head on the cup and a bubble sign. This article presents a case of a catastrophic failure of a ceramic/polyethylene bearing with destruction of the polyethylene liner and the metallic shell and protrusio of the nonfractured ceramic head through the metallic shell. To our knowledge this is the first description of extensive metallosis and subsequent radiograph bubble sign not presenting as a result of wear of a metal-on-metal articulation. At the time of revision surgery-Hydrocel TNT Monoblock 58 cup (Zimmer, Warsaw, Indiana), Wagner 265/14 stem (Zimmer), and a Co/Cr 28-mm head-copious metallic debris was seen both macroscopically and histologically, with the ceramic head protruding behind the metallic shell. Multiple factors may have been responsible for this failure including a thin polyethylene shell, a suboptimal locking mechanism, gamma in air sterilization for polyethylene, multiple screw-holes that reduce the contact surface between shell and polyethylene, the rough surface on the inside of the shell and non-articular wear at the metal polyethylene interface within the acetabular component and the high demands of this active young patient.
机译:一名有青少年类风湿关节炎病史的24岁患者接受了原发性非骨水泥性左全髋关节置换术(THA)。最初的THA包括一个带有3毫米厚聚乙烯衬里的Optifix 54杯,一个Optifix 4号茎(Smith&Nephew Richards,孟菲斯,田纳西州)和一个Biolox铝制32毫米股骨头。十四年后,X射线照片显示聚乙烯内衬的广泛磨损导致杯子上的陶瓷股骨头直接关节运动和磨损,并出现气泡迹象。本文介绍了一种陶瓷/聚乙烯轴承发生灾难性故障的情况,其中聚乙烯衬里和金属外壳遭到破坏,未破裂的陶瓷头穿过金属外壳而突出。据我们所知,这是首次广泛的金属化病的描述,随后由于金属对金属的关节磨损而没有出现放射线照相气泡迹象。在翻修手术时,从宏观和组织学角度观察到Hydrocel TNT Monoblock 58杯(齐默,华沙,印第安纳州),瓦格纳265/14茎(齐默)和Co / Cr 28毫米头部丰富的金属碎片。陶瓷头突出在金属壳后面。造成这种故障的原因可能有多种,包括聚乙烯外壳薄,锁紧机构欠佳,对聚乙烯进行空气灭菌时的伽玛,减少外壳与聚乙烯之间接触面的多个螺孔,外壳内部的粗糙表面髋臼组件内的金属聚乙烯界面处的非关节磨损以及该活跃年轻患者的高要求。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号