OBJECTIVES: This study was conducted to evaluate the accuracy of two electronic apex locators, the Sono Explorer Mark 2 Junior (Hakusui, Osaka, Japan) and Apit 2 (Osada, Tokyo, Japan) in detecting root perforations. The adequacy of radiographs for identifying root perforations was also assessed. STUDY DESIGN: Thirty-two extracted human teeth were perforated in the middle third of the root and embedded in alginate. Determination of all perforations were carried out with K-files no. 25 attached to the apex locators tested. Two radiographs were taken at two angulations after each electronic measurement. The actual location of the file tip in relation to the perforation was determined with a stereomicroscope. A total of 512 radiographs were evaluated to attempt to identify root perforations. RESULTS: The mean distance of the file tip from the external outline of the root surface was short for both instruments. A statistical difference (p < 0.05) was found between the two apex locators in dry canals or if saline solution was present. There was no significant difference between the two instruments in the presence of sodium hypochlorite. Evaluators radiographically identified 45% of the root perforations when located in buccal-lingual directions. CONCLUSION: Under the in vitro conditions of this study, both devices determined the location of the perforations in an acceptable clinical range short of the root surface. Radiographs were found to be less reliable in identification of perforation locations.
机译:目的:本研究旨在评估两种电子先头定位器,即Sono Explorer Mark 2 Junior(日本大阪市白水市)和Apit 2(日本大阪市大阪市)在检测根部穿孔方面的准确性。还评估了射线照片是否足以识别根部穿孔。研究设计:32颗拔出的人类牙齿在牙根的中间三分之一处穿孔,并嵌入藻酸盐中。所有孔眼的测定均使用编号为K的文件进行。 25个连接到顶点定位器上进行测试。每次电子测量后,在两个角度拍摄两张射线照片。使用立体显微镜确定锉刀尖端相对于穿孔的实际位置。总共对512张射线照片进行了评估,以试图确定根部穿孔。结果:两种工具的锉尖与牙根表面外轮廓的平均距离均较短。在干渠中或存在盐溶液的情况下,两个顶点定位器之间存在统计学差异(p <0.05)。在存在次氯酸钠的情况下,两种仪器之间没有显着差异。当位于颊舌方向时,评估人员通过射线照相确定根部穿孔的45%。结论:在这项研究的体外条件下,这两种装置均确定了穿孔的位置在可接受的临床范围内(不包括牙根表面)。发现射线照相在确定穿孔位置方面不太可靠。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号