Radiofrequency heating and magnetic field interactions of fixed partial dentures during 3-tesla magnetic resonance imaging

机译:固定局部义齿在3特斯拉磁共振成像过程中的射频加热和磁场相互作用

获取原文
获取原文并翻译 | 示例
           

摘要

Objective This study evaluated the heating and magnetic field interactions of fixed partial dentures in a 3-Tesla (3T) magnetic resonance imaging (MRI) environment. Study design Three substructure materials (Co-Cr, Ni-Cr, ZrO 2) were used to fabricate twelve 4-retained bridges and 12 crowns. Specimens were evaluated at 3T for radiofrequency heating and magnetic field interactions. One-way analysis of variance (ANOVA) test was used to compare continuous variables of temperature change. Significance was set at P .05. Translational attraction and torque values of specimens were also evaluated. Results None of the groups exhibited excessive heating (mean temperature change, 1.4 C), with maximum increase at the end of the T-1. Moreover, in all groups, only relatively minor magnetic field interactions that would not cause movement in situ were observed. Conclusion The study findings indicated that patients with fixed partial dentures (single crown or bridge) fabricated from Co-Cr, Ni-Cr, and zirconia substructures may safely undergo MRI at up to 3T.
机译:目的这项研究评估了在3-Tesla(3T)磁共振成像(MRI)环境中固定局部义齿的加热和磁场相互作用。研究设计使用三种下部结构材料(Co-Cr,Ni-Cr,ZrO 2)来制造十二个4座桥和12个冠。在3T下评估样品的射频加热和磁场相互作用。使用单向方差分析(ANOVA)测试来比较温度变化的连续变量。显着性设定为P <.05。还评估了样品的平移吸引力和扭矩值。结果所有组均未表现出过热(平均温度变化,<1.4 C),在T-1末期最大增加。此外,在所有组中,仅观察到不会引起原位移动的相对较小的磁场相互作用。结论研究结果表明,由Co-Cr,Ni-Cr和氧化锆亚结构制成的固定局部义齿(单冠或桥)的患者可以安全地接受3T的MRI检查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号