Intraobserver and interobserver reproducibility in the evaluation of optic disc stereometric parameters by Heidelberg Retina Tomograph.

机译:海德堡视网膜断层扫描仪在评估视盘立体参数时,观察者之间和观察者之间的可重复性。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To assess intraobserver and interobserver reproducibility of the measurement of stereometric parameters of the optic disc by means of the Heidelberg Retina Tomograph I (HRT). STUDY DESIGN: Observational study, with interobserver variability. PARTICIPANTS: Fifty-five volunteers (healthy subjects and patients with glaucoma). METHODS: HRT examination of the optic disc was repeated on 3 consecutive days on 1 eye of each of the 55 subjects. During each session, five single images were randomly acquired by two independent observers. One mean topography image (MTI), based on three single images, was then built at each session. For the intraobserver intraimage evaluation, the two observers traced their own contour line on one randomly chosen MTI. This procedure was repeated three times. For the intraobserver interimage and interobserver intra/interimage evaluations, the two observers traced their own contour line on the MTI of the first session, which was then automatically superimposed on the MTIs of the other two sessions. MAIN OUTCOME MEASURES: Reproducibility of the 12 stereometric parameters was calculated for each observer by means of the intraclass correlation coefficient (ICC). The expected range of variability between two independent evaluations was calculated by the scatter-plots of each test-retest difference versus their mean. The standard deviation of the mean test-retest score differences was used to describe the spread of score differences. RESULTS: The ICC ranged between 0.79 and 0.99 for intraobserver intraimage and between 0.56 and 1 for intraobserver interimage evaluation. The ICC ranged between 0.54 and 0.99 for interobserver intraimage and between 0.65 and 0.97 for the interobserver interimage evaluation. ICC was almost perfect to perfect for planimetric measures (0.81 < ICC < or = 1), substantial to almost perfect for volumetric and cup measures (0.61 < ICC < or = 0.99), and moderate to almost perfect for retinal nerve fiber layer related measures (0.41 < ICC < 0.99). The expected variability was low (95% confidence interval, < +/-9%). Interimage evaluation showed a higher variability than intraimage evaluation in both of interobserver (P = 0.012) and intraobserver evaluation (P = 0.028 and P = 0.031 for the two observers). CONCLUSIONS: Measurement of optic disc stereometric parameters by HRT is highly reproducible. However, the use of retinal nerve fiber layer-related parameters should be taken cautiously. The image acquisition-induced variability seems larger than the operator-induced variability.
机译:目的:通过海德堡视网膜断层扫描仪I(HRT)评估视盘立体参数测量值的观察者内部和观察者之间的可重复性。研究设计:观察性研究,观察者间存在差异。参与者:55名志愿者(健康受试者和青光眼患者)。方法:对55名受试者中的每只眼睛连续3天重复进行HRT视盘检查。在每个会话期间,由两个独立的观察者随机获取五张单张图像。然后在每个会话中建立一个基于三个单一图像的平均地形图像(MTI)。对于观察者内图像内评估,两个观察者在一个随机选择的MTI上绘制了自己的轮廓线。将该程序重复三遍。对于观察者内图像间和观察者间图像间/图像间评估,两位观察者在第一届会议的MTI上跟踪了自己的轮廓线,然后自动将其叠加在其他两次会议的MTI上。主要观察指标:通过组内相关系数(ICC)为每个观察者计算12个立体参数的可重复性。两次独立评估之间预期的变异性范围是通过每个重测差值相对其均值的散点图来计算的。平均重测分数差异的标准偏差用于描述分数差异的分布。结果:观察者内图像的ICC范围为0.79至0.99,观察者内图像评估的ICC范围为0.56至1。对于观察者间图像,ICC介于0.54至0.99之间;对于观察者间图像评估,ICC在0.65至0.97之间。 ICC对于平面测量几乎是完美至完美(0.81

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号