Stability of corneal polarization axis measurements for scanning laser polarimetry.

机译:用于扫描激光偏振法的角膜偏振轴测量的稳定性。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: Corneal polarization axis (CPA) has been reported to affect retardation measurements obtained with scanning laser polarimetry. The purpose of this investigation was to evaluate the longitudinal stability of CPA measurements. DESIGN: Prospective, noncomparative case series. PARTICIPANTS: Persons with normal corneas were enrolled; eyes with less than 1 year of follow-up from the initial CPA measurement were excluded. INTERVENTION: We constructed a noninvasive slit-lamp-mounted device incorporating two crossed linear polarizers and an optical retarder to measure the slow axis of corneal birefringence, or CPA. MAIN OUTCOME MEASURES: Corneal polarization axis measurements. RESULTS: Seventy-one eyes of 40 individuals (23 female, 17 male) were enrolled in this investigation (mean age, 42.9 +/- 13.6 years; range, 22-85 years). Initial CPA measurements (mean, 24.0 +/- 18.0 degrees nasally downward; range, 67 degrees downward to 13 degrees nasally upward) were strongly associated (R2 = 0.88; P < 0.0001) with repeat CPA measurements (mean, 20.9 +/- 14.6 degrees nasally downward; range, 59 degrees nasally downward to 14 degrees nasally upward). The mean change in CPA was 4.1 +/- 3.2 degrees (range, 0-13 degrees ). Corneal polarization axis stability was statistically associated with the mean (initial and repeat) CPA (R2 = 0.1; P = 0.009), but not associated with age (R2 = 0.0003; P = 0.9) or gender (R2 = 0.03; P = 0.2). CONCLUSIONS: Corneal polarization axis measurements have good 1-year stability. These data suggest that CPA should not contribute significantly to longitudinal measurements of retinal nerve fiber layer thickness obtained with scanning laser polarimetry.
机译:目的:据报道,角膜偏振轴(CPA)会影响通过扫描激光偏振仪获得的延迟测量。这项研究的目的是评估CPA测量的纵向稳定性。设计:前瞻性,非对比案例系列。参加者:角膜正常的人。从初始CPA测量中随访不到1年的眼睛被排除在外。干预措施:我们构建了一种无创裂隙灯安装设备,其中包括两个交叉的线性偏振器和一个光学延迟器,用于测量角膜双折射或CPA的慢轴。主要观察指标:角膜极化轴测量。结果:40名个体的71只眼(23名女性,17名男性)参与了这项研究(平均年龄42.9 +/- 13.6岁;范围22-85岁)。初始CPA测量值(平均,鼻翼向下24.0 +/- 18.0度;范围,下降67度到鼻翼向上13度)与重复CPA测量值(平均值,20.9 +/- 14.6)紧密相关(R2 = 0.88; P <0.0001)鼻下向下;度数范围:鼻下59度到鼻上14度)。 CPA的平均变化为4.1 +/- 3.2度(范围为0-13度)。角膜极化轴稳定性在统计学上与平均(初始和重复)CPA(R2 = 0.1; P = 0.009)相关,但与年龄(R2 = 0.0003; P = 0.9)或性别(R2 = 0.03; P = 0.2)不相关)。结论:角膜极化轴测量具有良好的1年稳定性。这些数据表明,CPA不应对通过扫描激光偏振法获得的视网膜神经纤维层厚度的纵向测量产生显着贡献。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号