Results of the calibration of Optical Time Domain Reflectometers (OTDR) according to IEC-proposals will be presented. The linearisation of the power scale was performed by the ''Power reduction method-long fiber alternative (PRM-LFA)'' and the ''Loss calibration with fiber standard''. The test procedures as well as the correction of the displayed OTDR attenuation value are described, It will be shown that the PRM-LFA is the superior technique with respect to performance and accuracy, After calibration and linearisation the uncertainty of the attenuation coefficient can be reduced down to 0.002 dB/km. [References: 5]
展开▼
机译:将介绍根据IEC提案的光时域反射仪(OTDR)的校准结果。功率刻度的线性化是通过“功率降低方法-长光纤替代(PRM-LFA)”和“光纤标准损耗校准”来进行的。描述了测试程序以及显示的OTDR衰减值的校正,结果表明PRM-LFA在性能和准确性方面是上乘的技术,经过校准和线性化后,可以降低衰减系数的不确定性低至0.002 dB / km。 [参考:5]