An 840-nm wavelength spectral domain optical coherence tomography (SD-OCT) with prolonged scan depth was developed and mounted onto a conventional slit lamp for imaging the anterior segment of the eye. X-Y cross aiming was applied to align the SD-OCT scanning position during imaging. An internal fixation target displayed on a miniature LCD monitor was provided. The SD-OCT instrument had an axial resolution of 6 um and a prolonged scan depth of 7.2 mm. High-quality SD-OCT images, consisting of 2,048 X 2,048 pixels, were acquired of the entire anterior chamber and entire crystalline lens from a healthy subject. The entire cornea, anterior chamber angle, limbus, and iris were clearly visible. Additionally, the internal structure of crystalline lens, including anterior and posterior surfaces of the crystalline lens, capsule, nucleus, and cortex, were clearly imaged with the instrument. The system was also tested in imaging accommodation of the same eye, demonstrating the feasibility of the novel approach for evaluating presbyopia/accommodation.
机译:开发了具有延长扫描深度的840-nm波长光谱域光学相干断层扫描(SD-OCT),并将其安装在用于成像眼前节的常规裂隙灯上。 X-Y交叉瞄准用于在成像过程中对齐SD-OCT扫描位置。提供了在微型LCD监视器上显示的内部注视目标。 SD-OCT仪器的轴向分辨率为6 um,延长的扫描深度为7.2 mm。从健康受试者的整个前房和整个晶状体中获取高质量的SD-OCT图像,该图像由2,048 X 2,048像素组成。清晰可见整个角膜,前房角,角膜缘和虹膜。另外,用该仪器清晰地成像了晶状体的内部结构,包括晶状体的前表面和后表面,囊,核和皮质。该系统还在同一只眼睛的成像适应性中进行了测试,证明了该新方法用于评估老花眼/适应性的可行性。