Ion selective membrane electrodes for determination of benoxinate hydrochloride in pure form and in drug product

机译:离子选择性膜电极,用于测定纯净形式和药品中的苯甲酸酯盐酸盐

获取原文
获取原文并翻译 | 示例
       

摘要

This paper presents a comparative study between four sensors constructed to determine benoxinate hydrochloride (BX) in the presence of its hydrolysis induced degradation product using different ion association complexes and plasticizers. Precipitation based technique was used for sensors fabrication. The BX complexes with the cationic exchangers; BX-reinikate, BX- tetraphenylborate, BX-phosphotungestate, and BX- tetrakis were obtained in situ by soaking the PVC membranes in 1 × 10~(-2) BX solution. Dioctylphthalate and nitrophenyl octyl ether were used as solvent mediators. The proposed sensors showed fast, stable Nernstian responses across a relatively wide BX concentration range (5×l0~(-5) to 10~(-1)M) in the pH range of 4-6. The suggested sensors could be used for several weeks without any measurable change in sensitivity. They displayed good selectivity for BX in presence of its degradation product, common inorganic and organic species. The proposed sensors were successfully applied for the determination of BX in pure powder form and eye drops where good recoveries were obtained.
机译:本文介绍了四种传感器之间的比较研究,该传感器用于使用不同的离子缔合配合物和增塑剂,在水解诱导降解产物存在下,测定苯甲酸钠盐酸盐(BX)。基于沉淀的技术被用于传感器的制造。 BX与阳离子交换剂形成复合物;通过将PVC膜浸泡在1×10〜(-2)BX溶液中原位获得BX-reinikate,BX-四苯基硼酸酯,BX-磷钨酸酯和BX-四烷基酯。邻苯二甲酸二辛酯和硝基苯基辛基醚用作溶剂介体。提出的传感器在4-6的pH范围内,在相对宽的BX浓度范围(5×10〜(-5)至10〜(-1)M)中显示出快速,稳定的能斯特响应。建议的传感器可以使用数周,而灵敏度没有任何可测量的变化。它们在降解产物,常见的无机和有机物质的存在下对BX具有良好的选择性。所提出的传感器已成功应用于纯粉末形式和滴眼液中BX的测定,并获得了良好的回收率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号