首页> 外文期刊>Rheumatology >Automated counting of white blood cells in synovial fluid.
【24h】

Automated counting of white blood cells in synovial fluid.

机译:滑膜液中白细胞的自动计数。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVES: To evaluate the performance of automated leucocyte (white blood cell; WBC) counting by comparison with manual counting. METHODS: The number of WBC was determined in heparinized synovial fluid samples by the use of (i) a standard urine cytometer (Kova) and a microscope (reference method) and (ii) a haematology analyser (Sysmex XE-2100; WBC/BASO and DIFF channels). Imprecision within and between days was determined by replicate analysis of a low (level A; WBC approximately 0.560 x 10(9)/l) and a high (level B; WBC approximately 1.081 x 10(9)/l) dedicated synovial fluid control (Quantimetrix). RESULTS: The WBC count of the DIFF channel was highly correlated with the WBC count of the microscopic reference method (r = 0.99; WBC analyser = 0.870 x WBC reference method + 0.413). In contrast, no agreement existed between WBC counts generated by the WBC/BASO channel of the analyser and the reference method (r = 0.52; WBC analyser = 0.008 x WBC reference method + 0.079). Within-day imprecision (4-7%)and between-day imprecision (10%) of the haematology analyser were smaller than the within-day imprecision (12%) and the between-day imprecision (20-22%) of the manual reference method. For manual counting, inter-observer coefficients of variation were 35.9% (control level A) and 21.0% (control level B). CONCLUSIONS: The WBC count in synovial fluid can be reliably determined using the DIFF channel of the Sysmex XE-2100. Automated counting of WBC in synovial fluid offers more precise and faster results than manual counting.
机译:目的:通过与手动计数比较来评估自动白细胞计数(白细胞; WBC)的性能。方法:通过使用(i)标准尿细胞计数仪(Kova)和显微镜(参考方法)和(ii)血液分析仪(Sysmex XE-2100; WBC / BASO)确定肝素化滑液样品中的白细胞数量和DIFF频道)。通过重复分析低(A级; WBC约0.560 x 10(9)/ l)和高(B级; WBC约1.081 x 10(9)/ l)专用滑液进行重复分析,确定几天之内和之间的不准确性(Quantimetrix)。结果:DIFF通道的WBC计数与微观参考方法的WBC计数高度相关(r = 0.99; WBC分析仪= 0.870 x WBC参考方法+ 0.413)。相反,分析仪的WBC / BASO通道生成的WBC计数与参考方法之间没有一致性(r = 0.52; WBC分析器= 0.008 x WBC参考方法+ 0.079)。血液分析仪的日内不准确度(4-7%)和日间不准确度(10%)小于手册的日内不准确度(12%)和日间不准确度(20-22%)参考方法。对于手动计数,观察者之间的变异系数为35.9%(对照水平A)和21.0%(对照水平B)。结论:滑液中的白细胞计数可以通过Sysmex XE-2100的DIFF通道可靠地确定。与手动计数相比,滑液中WBC的自动计数可提供更精确,更快速的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号