首页> 外文期刊>Radiation Physics and Chemistry >An international intercomparison on'self-calibrated'alanine EPR dosimeters
【24h】

An international intercomparison on'self-calibrated'alanine EPR dosimeters

机译:“自校准”丙氨酸EPR剂量计的国际比较

获取原文
获取原文并翻译 | 示例
           

摘要

The results,obtained by six independent electron paramagnetic resonance(EPR)laboratories,of the dose response coefficients(K_(dr))of"self-calibrated"solid-state EPR dosimeters containing alanine as a radiation-sensitive material and Mn~(2+)/MgO as an internal reference material,are reported.The intercomparison trial was divided into three steps.It started with the distribution of dosimeters among the participating EPR laboratories with the purpose of irradiating them with known doses of gamma-rays and to estimate the K_(dr).The percentage standard deviation(PSD)of the K_(dr)obtained at individual labs was in the range of 1.4-4.6%.The interlaboratory PSD of the K_(dr)was 8%,primarily pointing to variations in irradiation procedures and EPR spectrometer settings.Further investigations showed that the main source of the interlaboratory PSD is differences in the calibrations of irradiators and settings of EPR acquisition parameters.In order to provide reproducible estimates of the K_(dr),low microwave power and modulation amplitude using a combination of sweep time and time constant that gives a distortion-free EPR spectrum should be utilized.In the third step following such a procedure,measuring the same irradiated alanine dosimeter at the respective laboratories,spectrometers(12 instruments of 6 different models and 3 producers)and 10 operators gave an interlaboratory PSD of 3.1%.In conclusion,EPR dosimetry using"self-calibrated"alanine dosimeters may be used as a secondary standard,although a careful calibration of the EPR spectrometer must be performed in order to further reduce the uncertainty.
机译:通过六个独立的电子顺磁共振实验室获得的结果,以丙氨酸作为辐射敏感材料的“自校准”固态EPR剂量计的剂量响应系数(K_(dr))为Mn〜(2对比试验分为三个步骤。首先,在参与的EPR实验室中分配剂量计,目的是用已知剂量的γ射线辐照剂量计并估算各个实验室获得的K_(dr)的百分比标准偏差(PSD)在1.4-4.6%的范围内.K_(dr)的实验室间PSD为8%,主要指出了差异进一步的研究表明,实验室间PSD的主要来源是辐照器的校准和EPR采集参数设置的差异,以便提供可重现的K_(dr),l应当利用扫频时间和时间常数相结合的低功率和调制幅度来产生无失真的EPR光谱。在此步骤的第三步中,在各个实验室,光谱仪上测量相同的辐照丙氨酸剂量计(12 6个不同模型和3个生产商的仪器和10个操作员的实验室间PSD为3.1%。总之,尽管必须仔细校准EPR光谱仪,但使用“自校准”丙氨酸剂量计的EPR剂量学可以用作次要标准。为了进一步减少不确定性而进行。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号