Two-dimensional thermoluminescence dosimetry system for proton beam quality assurance

机译:用于质子束质量保证的二维热释光剂量测定系统

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The response of a large-area two-dimensional (2D) thermoluminescence (TL) dosimetry system with 20 x 20 cm(2) TL LiF:Mg,Cu,P foils, developed at the Institute of Nuclear Physics PAN was studied for doses of therapeutic proton beams with energies of 80 MeV, 150 MeV and 225 MeV. The one-hit detector model fitted to measured dose response yielded characteristic doses of 227 Gy, 209 Gy and 203 Gy respectively. The system was applied to investigate geometrical parameters of spott produced by proton pencil beams at the Bronowice Cyclotron Centre IFJ PAN. Mean spot sizes measured with TL foils were compared with mean spot sizes calculated for Gafchromic (R) films. It was shown that the 2D TL system was capable to measure the spot size up to peak doses of 20 Gy without additional correction for dose response. For measurements of the spot size using Gafchromic (R) EBT3 films an additional calibration must be applied to correct for the non-linear response for doses exceeding 1 Gy. (C) 2015 Elsevier Ltd. All rights reserved.
机译:研究了由PAN核物理研究所开发的具有20 x 20 cm(2)TL LiF:Mg,Cu,P箔的大面积二维(2D)热致发光(TL)剂量系统的响应。能量为80 MeV,150 MeV和225 MeV的治疗性质子束。符合所测剂量响应的一击检测器模型分别产生了227 Gy,209 Gy和203 Gy的特征剂量。该系统用于研究布罗诺维斯回旋加速器中心IFJ PAN质子笔形束产生的斑点的几何参数。将用TL箔测量的平均光斑尺寸与为Gafchromic(R)膜计算的平均光斑尺寸进行比较。结果表明,二维TL系统能够测量最大20 Gy剂量的斑点大小,而无需对剂量响应进行其他校正。对于使用Gafchromic(R)EBT3膜的光斑尺寸测量,必须进行额外的校准,以校正剂量超过1 Gy时的非线性响应。 (C)2015 Elsevier Ltd.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号